Device Verification
Verify the device function
Step 1:
Set up the system. Refer to the Aulisa sensor module(s) Instructions for Use
for setting up instructions.
Step 2:
Make sure Aulisa sensor module(s) is worn on or attached to the right place
firmly.
Step 3:
Verify that the Bluetooth connection status icon on the screen is blue and
the status indicator on Aulisa sensor module(s) is blinking green.
Step 4:
Verify that the vital sign readings are displayed.
Verify the alarm function BEFORE each use
Step 1:
Set up the system. Refer to the Aulisa sensor module(s) Instructions for Use
for setting up instructions.
Step 2:
Ensure the Bluetooth connection is established. (See “Device Pairing”
section.)
Step 3:
Set the alarm limit lower or higher than the displayed value.
Step 4:
Verify that an alarm message is displayed and that an alarm audio is
generated. (See “Troubleshooting” section if an alarm message and audio
signal is not generated.)
Step 5:
Press on the "PAUSE AUDIO" button to temporarily silence for 2 minutes.
Step 6:
After alarm signal is regenerated, press on the "AUDIO OFF" button to
silence permanently the alarm signal.
Step 7:
Tap on “RESTORE DEFAULTS” to restore alarm limits to manufacturer
configured values before use.