INSTRUCTIONS FOR USE
The instructions below describe how to operate the Controller. For information about Warming Devices and
accessories, refer to the User Manual provided with each item.
1.
Insert the Controller power plug into a properly grounded hospital-grade electrical outlet. Turn on the
Controller via the rocker switch at rear.
WARNING: To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth grounding.
Note: The Controller is grounded and should not be attached to un-grounded tables intended for use with a
hyfrecator or equivalent devices.
2.
Position and secure the Device (e.g., Warming Blanket, Warming Mattress) following instructions in the User
Manual for each device.
3.
Insert the device’s connecting cable into the proper port on the Controller.
Table 1.
WC77 Multi Port
WC71 Single Port Port Color
Device
A, B, C, D
A
Yellow
Warming Blanket, Clinician Vest
M
Blue
Warming Mattress
T
probe
(In)
T
probe
(In)
Silver
Patient Temperature Probe (3.5mm TSR jack)
T
out
T
out
Silver
3.5mm TR Jack (YSI 400 compatible output)
Figure 2: Controller Ports
Port A on front of
Controller
Ports B, C, D and M on
back upper left of the
Controller
(Multiport only)
Back of the Controller,
lower left
T
TruCore
(Not Currently
Available), T
probe
and T
out
on
back of Controller (Single Port
and Multiport)
Note: When the connecting cable is inserted into the Controller, an audible sound indicates that the control sensor
and over-temperature thermistor are present and functioning properly. The Port becomes available on the
touchscreen.