AA222
Operation Manual
Page 85
Paper Rolls:
Width:
112 (+/- 0.5) mm
Diameter:
45 (+2.0) mm
Basis weight:
58.5 +/- 3g/m2
Thickness:
62 +/- 3 u
Tensile:
Min. 3.2 kN/m
Impedance Measuring System
Probe tone:
Frequency :
226 Hz.
Level :
85 dB SPL with AGC, assuring
constant level at different volumes.
Air Pressure
Control:
Automatic.
Indicator:
Measured value is displayed on the
graphical display.
Range:
-600 to +300 daPa.
Pressure Limitations:
-800 daPa and +600 daPa.
Pressure Change Rate:
Minimum (50 daPa/s), medium,
maximum or automatic with
minimum speed at compliance
peak. Selectable in the setup.
Compliance
Range:
0.1 to 6.0 ml (Ear volume: 0.1 to 8.0
ml). No difference is measured in the
recorded values between static and
dynamic measuring mode.
Types
Tympanometry:
Automatic, where the start and stop
pressure can be user-programmed
from the setup menu
.
Eustachian Tube
Function:
Williams test (automatic
function).