AA222
Operation Manual
Page 78
Picture 9:
After unscrewing the probe cap it is possible to detach the probe tip
with the small acoustic and air pressure channels from the
transducer house. Out of the probe tip you can take temporarily
remove the sealing gasket.
Picture 10:
Probe cap Sealing gasket Transducer House
Probe Tip with the small acoustic
and air pressure channels
The cleaning of the acoustic and air pressure channels of the probe
tip must be performed by means of the cleaning wire (nylon wire)
which is part of the cleaning tool that can be found in the Ear tips
Assortment provided with the AA222.
When cleaning the acoustic and air pressure channels of the probe
tip the cleaning wire must be inserted from the back of the probe tip
according to picture 11: