![Audio Precision Portable One Plus Access User Manual Download Page 159](http://html1.mh-extra.com/html/audio-precision/portable-one-plus-access/portable-one-plus-access_user-manual_3008598159.webp)
GenLoad Function
Measurement Range
<1
Ω
to 20 k
Ω
Accuracy
±
[5 % + 0.5
Ω
] for readings <1 k
Ω
Degrades rapidly above 1 k
Ω
or with
reactive loads
Frequency Range
20 Hz to 20 kHz
Test Voltage
200 mV default. Usable from 10 mV to
generator maximum
Wow & Flutter Function
Test Signal Compatibility
2.80 kHz to 3.35 kHz
Measurement Range
<0.005% to 3% (single range)
Accuracy (4 Hz)
±
(5% of r 0.002%)
Detection Modes
IEC/DIN (quasi-peak per IEC-386),
NAB (average),
JIS (per JIS 5551)
Response Selections
Weighted4 Hz band
pass per IEC/DIN/NAB
Unweighted
0.5 Hz-200 Hz
Residual W+F
Weighted
≤
0.005%
Unweighted
≤
0.01%
Minimum Input
25 mV
Settling Time
IEC/DIN or NAB
Typically 3 to 6 seconds
JIS
Typically 15 to 20 seconds
G
S
p
ec
ificat
ions
Appendix G - Specifications
Portable One Plus Access User's Manual
G-7
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com