The upper right corner ofthe display shows the current firmware
version. This is the version ofthe internal program that runs the
Portable One Plus Access, and applies to all features, not just the self
test.
The lower left soft key will run the self test. Once the self test begins,
the center section ofthe display (where the generator parameters are
normally displayed) will show the number and name ofthe test being
run. The upper center section ofthe display will show ‘RUNNING’ to
indicate that the selftest is in progress.
Ifan error is encountered, the selftest will pause. The upper center
section ofthe display will show ‘ERROR’ and the value ofthe data that
caused the failure. You may want to write this data down so that you
may provide it to support personnel. Ifyou have a printer connected,
you may want to press ‘PRINT’ instead, to print the entire display.
The lower left soft key will change to ‘CONTINUE’. When you are
done observing the failure, you may press this key to continue with the
selftest. Alternatively, you may press the lower right soft key (‘QUIT’)
to stop the selftest. Pressing QUIT at any time will return you
immediately to the Main Panel.
When the selftest is finished, you will be returned to the SelfTest
panel. The upper center section ofthe display will read ‘COMPLETE’
and the generator section ofthe display will show ‘PASSED’ or
‘FAILED’ depending on the results ofthe test.
ATTENTION!
Always disconnect all cables from the generator
outputs before running the self test. Connections to the
generator outputs may cause the self test to fail, and may
damage the device-under-test.