Characteristic
Specifications
Supplemental Information
APx1701 Transducer Test Interface: Specifications
11
Input Impedance
4
Balanced, each side to ground
≈
20 k
Ω
|| 240 pF–300 pF
Unbalanced
≈
20 k
Ω
|| 240 pF–300 pF
Input AC Coupling
time constant
≈
0.20 s;
–3 dB is
≈
0.8 Hz
Switching Configuration
Microphone Output
(Pass Thru) 1
Switchable from among microphone bal-
anced inputs 1 and 2, or microphone
unbalanced inputs 1 and 2. Also switch-
able to read the DC output of either
unbalanced microphone input.
Microphone Output
(pass Thru) 2
Switchable from among microphone
balanced input 2 or microphone
unbalanced input 2 only.
Residual Noise
Determined by the audio analyzer; the
inputs are passively switched.
Input Crosstalk
(R
S
≤
600
Ω
)
≤
(–80 dB + 1
μ
V) to 20 kHz
OUTPUT CURRENT SENSE (MONITOR)
Sensitivity
0.100 V per Amp, ±0.50%
0.10
Ω
current sense resistor is in series
with the – (minus) Output terminal of
each amplifier.
Switching Configuration
Power amplifier output 1 or 2. Current
sense resistor remains in circuit at all
times even when not selected.
Summary of Contents for APx1701 Series
Page 1: ......
Page 2: ......
Page 6: ......
Page 8: ...Chapter 1 Table of Contents APx1701 Transducer Test Interface Table of Contents...
Page 14: ...S curit vi APx1701 Transducer Test Interface S curit...
Page 18: ...Seguridad x APx1701 Transducer Test Interface Seguridad...
Page 26: ...8 Abbreviations Terms and Symbols...
Page 42: ...Chapter User s Guide 24 APx1701 Transducer Test Interface User s Guide...
Page 43: ......
Page 44: ......