18 Setting up and starting up
4. If the device is not damaged: place the device on a safe, level surface.
3.3 Connecting to the mains power supply
1. Check whether the voltage and frequency data listed on the device correspond to
the values of the mains power supply.
2. Connect the device to the mains.
3. Secure the power cable against accidental removal by using the safety clamp.
For surgical procedures, we recommend additionally connecting the device via the
connection to the potential equalisation of the examination room.
3.4 Connecting the secretion canister system and hoses
1. The anti-foaming device must be placed on the
fill-level sensor for strongly foaming secretion.
2. Slide the secretion canister lid with the release
button pointing forward onto the secretion
canister.
Make sure that the lid rim is under the bead of
the secretion canister. This seals the secretion
canister tightly and the desired vacuum can be
built up in the secretion canister.
3. Press the locking handle downwards until it
clicks into place.
4. Hang the secretion canister in the left or right
support for the canister system.
Using the support for the canister system for
purposes other than what it was designed for can
cause malfunctions.
1. Push the double hose connector into the
canister lid twisting gently.
» The double hose connector locks into place.