9.1 Basic information
● Carry out a visual inspection of the unit prior to each use
including hoses, collection canister and connection cable.
Da-
maged cables and hoses must be replaced immediately.
● Maintenance or opening and repair of the ATMOS A / C 161
/ 261 Aspirator (with the exception of the cleaning work
described in these operating instructions) may only be carried
out by ATMOS or a specialist authorised by ATMOS. In this
case, attention should be paid to the protective technical and
hygiene measures, the notes on safety plus the descriptions
in the servicing instructions for the ATMOS
®
C161.
● For repair, this device can be returned to ATMOS.
● Before returning the device for repair, clean and afterwards
disinfect all secretion canister parts and hose parts. The
device´s surface also has to be disinfected.
● ATMOS cannot guarantee perfect functioning neither will it be
liable for damage to people or property if:
• Any non-original ATMOS parts are used,
• the user instructions given in this manual are not
followed exactly or are disregarded,
• assembly, resetting, alterations, extensions and
repairs are not carried out by people authorised by
ATMOS.
● No warranty rights shall exist in the event of damage or
failure caused by the use of non-ATMOS accessories or non-
ATMOS consumables.
● In order to protect the user, the ATMOS
®
A / C 161 Batte-
ry must be in good condition with regard to technics and
hygiene prior to passing on. If there is a change of patient
or ownership then the device must be reprocessed acc. to
Medical Devices Operator Ordinance, Medical Devices Act
and BVMed guidelines. An oversucked (contaminated) device
must be repaired by the manufacturer, by a certifi ed ATMOS
partner or an ATMOS authorized, specialized dealer.
● Pay attention to regulations and instructions valid for the
respective application range.
9.0 Maintenance
condensate
collector for a
quick view control
of a possible
contamination
9.2 Reprocessing
The way the suction device is used determines its
reliability and safety. These hygiene measures descri-
bed in the last chapter are indispensable for protecting
the patient and the user and for maintaining a safe and
reliable suction device.
When you can ensure the device was not over-
sucked then perform a reprocessing acc. to Medical
Devices Operator Ordinance, Medical Devices Act
and BVMed guidelines. The reprocessing consists
of cleaning, surface disinfection as well as the
exchange of consumables. An ATMOS set of con-
sumables is available for this purpose.
If you cannot exclude that the device was over-
sucked then the device must be repaired by the
manufacturer, by a certifi ed ATMOS partner or an
ATMOS authorized, specialized dealer. Subsequent-
ly the device may be operated again.
Valid for ATMOS
®
C 161 / C 261 Aspirator
How can one realise that the suction device is con-
taminate?
Perform a visual inspection of the condesate collector
(at the device,
14
). In case of soiled or damped
from the condensate collectore, the device is oversu-
cked and therefore contaminated.
In case you realise this incidents, it is
necessary to have the device repaired by ATMOS or by
a certifi ed ATMOS Partner.
If any reservations exist regarding the hygiene condition
of the device, please send the device to ATMOS or a cert
fi ed partner for inspection.