6.3.2
Xenon lamp installation
1.
Fig. 17:
Insert the xenon lamp
2
into the lamp socket
9
and carefully push down all the way to the stop.
2. Insert the filter lantern
3
into the exterior cylinder
8
.
Make sure the filter lantern is properly seated in the
cylinder base ring.
CAUTION - Overheating!
Aside from filtering light, the light filters
also act as heat absorbers.
If tests are run without light filters, the in-
strument may overheat, resulting in damage
to the samples.
Only ignite the xenon lamp if light filters
are installed.
3. Place the spacer tube
5
onto the flange tube
7
.
4. Make sure the ground wire
6
is screwed to the flange
tube.
5. Close the exhaust cover with the connecting tube
4
.
6. Reinstall the connecting piece
1
for the exhaust duct
or technical ventilation system.
6
Start-up
Fig. 17