6.3.1
Removing the xenon lamp
1.
Fig. 16:
Remove the connector
1
for the exhaust
duct or technical ventilation.
2. Flip the exhaust cover with the connecting piece
4
up.
3. Lift the spacer tube
5
off the flange tube
7
and put it
aside. The spacer tube is merely pushed onto the
flange tube, but is connected to it by the ground wire
6
.
4. Pull the filter lantern
3
up and out of the top of the
exhaust duct.
5. Carefully grasp the used xenon lamp
2
at the upper
arc and pull it up and out of the lamp socket.
6
Start-up
Fig. 16