4.7
Measure Menu [MEASURE]
The default menu for the system is the display-only menu MEASURE. This menu
is a display-only measurement menu, and has no changeable list items. When
left alone, the instrument will automatically return to this menu after
approximately 30 minutes. While in the default menu, the UP arrow allows the
user to scroll through the secondary variables on the lower line of the display. A
brief description of the fields in the basic transmitter version is as follows:
TRANSMITTER MEAS SCREENS:
25.7C
Temperature display.Can be displayed in C or F, depending
on user selection. A small “m” on the left side of the screen
indicates the transmitter has automatically jumped to a
manual 25C setting due to a failure with the temperature
signal input.
Iout#1
Instrument output #1
Iout #2
Instrument output #2
Slope = 100%
Sensor calibration slope (updated after successful calibration
has been completed)
TDS = 200 mgL
Total Dissolved Solids (TDS)
.
Displays TDS of the process.
Q45CT Vx.xx
Transmitter software version number.
Note: A display test (all segments ON) can be actuated by pressing and
holding the ENTER key while viewing the model/version number on
the lower line of the display.