| Auto Wash Pro Manual | v1.2
Calibrating the Peristaltic Pumps
Press the Shift key
then press the Enter key
to enter the setup menu.
Press the 2 key
until the Pump Calibration menu item is reached.
Press the Enter key
to calibrate the peristaltic pumps.
Press the Up or Down arrow keys
to choose which peristaltic pump to
calibrate.
Put the pipe from the selected peristaltic pump into a measuring beaker / jug to collect the pumped
chemical. The measuring beaker / jug should be chemical resistant.
Press the Auto key
to run the selected peristaltic pump and measure the collected chemical.
Press the Enter key
to enter the measured amount of chemical.
Enter the value using the number keys
then press the Enter key
to
save the measured value.
Press the Cancel key
to return to the Setup menu.
Shift
Enter
Setup
Enter
Setup
pump calibration
pump 1 800ml
or
Auto
pump 1 ____ml
Enter