25 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Example: bit number 13 (Program name) activated to 1, will place to "2000h" the value in the first
word.
2000h is equivalent to 8192 in decimal and 0010000000000000 in binary.
In the Modbus frame, the words will follow each other: word 1 + word 2 + …..
+ word n.
Word
Bit n°
Mask
D.A. address
Meaning
Menu
Hexa
Dec
Read
Write
7
96
0001
1
249B
649B
Buzzer function.
+Funct
97
0002
2
249E
649E
Long test (x100) function.
+Funct
98
0004
4
249C
649C
Permanent blowing.
Conf
99
0008
8
24A4
64A4
Sealed Diff component function.
+Funct
100
0010
16
24A5
64A5
Test or Ref Mode.
+Funct
101
0020
32
24B9
64B9
Display optional.
+Funct
102
0040
64
24B6
64B6
Pressure Drop.
+Funct
103
0080
128
24B7
64B7
Pressure correction (≥v1.400)
.
+Funct
104
0100
256
24B8
64B8
Standard conditions (≥v1.400)
.
+Funct
105
0200
512
248F
648F
Ref No Dump.
+Funct
106
0400
1024
24BA
64BA
Auto Vol.
+Funct
107
0800
2048
24BB
64BB
Offset.
+Funct
108
1000
4096
24BC
64BC
Old Flow Calculation.
+Funct
109
2000
8192
Reserved.
110
4000
16384
24BE
64BE
Auto Selection Prog.
Conf
111
8000
32768
24BF
64BF
Save Volume Selection.
Conf