ERRORS AND FAULTS
See the manuals CDROM for further information.
1. ERROR MESSAGES
The
ATEQ F5800
can display error messages if there are operational problems.
Reference
fault /
Test
fault / Pressure in excess of the
full scale
/ Error on the
differential sensor
/ Pressure in
excess of the max. threshold
/ Pressure
below the
min. threshold
/
ATR
fault / fault or drift
CAL
/
Valve
commutation
fault /
PROG
Error
/
Inappropriate size
for the selected unit of pressure /
Sealed component
learning error /
Sealed component
error /
Sealed components
large leak error /
Auto-test
fault /
2. IN CASE OF OPERATION DOUBT
If a test machine begins to detect too many fail parts (more than three consecutively), it
is advisable to carry out a
check on the whole unit.
The quality of the manufacture and
operation of the leak detector should be the last things considered.
There is a possibility that the seals may be cut by shavings or worn by repetitive
squashing. This can be prevented by regular servicing and replacement of the seals.
If all the other checks do not resolve the problem, the instrument’s circuit may be
checked.
ATEQ
does not accept any liability in regard to calibrations and settings to its
instruments which are not carried out by its own personnel.