2. MEASUREMENTS CHARACTERISTICS
2.1. F
LOW MEASUREMENT RANGES
Flow
ranges
Standard
Option (*)
Maximum
resolution
4 ml/min
2% RV + 0,01 ml/min
1% RV + 0,01 ml/min
0,001 ml/min
Accuracy in
conditions
defined as
ambient
No lin
Hyst
Repeata T° drift
+ Long time drift (90
days)
40 ml/min
2% RV + 0,1 ml/min
1% RV + 0,1 ml/min
0,01 ml/min
4 ml/min
2,5% RV + 0,01 ml/min
2% RV + 0,01 ml/min
0,001 ml/min
Accuracy in
conditions
defined as
standard
(1013 hPa -
20°C)
No lin
Hyst
Repeata T° drift
+ Long time drift (90
days)
40 ml/min
2,5% RV + 0,1 ml/min
2% RV + 0,1 ml/min
0,01 ml/min
RV = Read value.
(*) Instrument adjusted in our metrology laboratory with COFRAC agreement.
2.2. A
TMOSPHERIC PRESSURE AND TEMPERATURE INDICATORS
Minimum
Maximum
Accuracy
(20°C)
Atmospheric pressure
850 hPa
1050 hPa
+/- 200 Pa
Temperature (internal)
0° C
50° C
+/- 1,5%
Note
: the zero offset linked to the temperature or the position can be take over by an
auto-zero. It is however advised to make an auto-zero before a measurement.
Recommendation of use:
according to some use's conditions, the instrument can put
to two hours of time of heating before complete stabilization.
Summary of Contents for CDF60
Page 1: ...USER MANUAL ATEQ CDF60 Version 1 0 Reference UM 23900B U...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 20: ...Chapter 2 User Interfaces UM 23900B U User manual ATEQ CDF60 Page 10 40...
Page 40: ...Chapter 6 Error messages UM 23900B U User manual ATEQ CDF60 Page 30 40...
Page 46: ...Chapter 7 PC installation UM 23900B U User manual ATEQ CDF60 Page 36 40...
Page 50: ...Index UM 23900B U User manual ATEQ CDF60 Page 40 40...
Page 51: ......