Atal ATC-06 Instruction Manual Download Page 6

6

Low battery voltage is indicated on the display with blinking reading "BAT". It is necessary to 

replace it with new one as soon as possible. Battery is located under small cover on the instrument 
lower side. It is absolutely necessary to replace battery with instrument switched OFF, otherwise 
setting of d.REF. and AUDI (from menu selections) and data in memory HOLD, MIN and MAX 
will  be  lost.  For  the  same  reason  do  not  disconnect  the  battery  for  longer  than  1 minute  even  if 
instrument is switched OFF. If it happens (or if battery is totally discharged), it is necessary to set 
again  in  appropriate  menu  selection  LCD  refreshment  mode  (d.REF.),  alarm  audio    indication 
(AUDI) and clear the minimum and maximum memory (CLR).

ATAL instruments passed the following electromagnetic compatibility (EMC) tests:

radiation: EN 55022

class B 

immunity: EN 61000-4-2

(levels 4/8 kV, class A)

EN 61000-4-3

(intensity of electromagnetic field 3 V/m, class B) 

EN 61000-4-4

(levels 1/0,5 kV, class A)

EN 61000-4-6

(intensity of electromagnetic field 3 V/m, class B) 

EN 61000-4-11

(class A)

EN 61000-4-5

(class A)

Summary of Contents for ATC-06

Page 1: ...ATAL ATC 06 Digital thermometer hygrometer Instruction manual ...

Page 2: ...ttery 9V or ac dc adapter 12V with NiCd accumulator 9V Average current consumption 0 15 to 0 7 mA depending on operation mode Typical life of Zinc Chloride battery 4 months Typical life of Alkali Mangan battery 6 months Operation conditions Ambient temperature range 10 to 60 C Ambient relative humidity range 5 to 95 RH Outer characteristics in accordance with EN 33 2000 3 normal environment with c...

Page 3: ...nt anytime after switching ON It is possible to enter this mode from other modes by pressing or by repeating pressing of MENU key Temperature in C is displayed on the upper LCD line and relative humidity in RH is displayed on the lower LCD line Press DISPLAY key to display other readings temperature in C on upper line and dew point temperature on lower line C DP Function HOLD storing of actual mea...

Page 4: ...NTER key to display values stored in the HOLD memory Press DISPLAY key to display other stored values dew point temperature Press MENU key to leave this mode Battery voltage of partially loaded battery is displayed This value illustrates battery condition Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode ...

Page 5: ... the arrow key UP to make value increase fast Press and hold the arrow key DOWN to make value decrease fast Release the arrow key and press ENTER to confirm new limit Reading ALHI indicates adjusted upper limit of the same alarm here air temperature Set the desired value in the same way as in above lower limit If needed it is possible to get back to lower limit setting of the same alarm by pressin...

Page 6: ...battery for longer than 1 minute even if instrument is switched OFF If it happens or if battery is totally discharged it is necessary to set again in appropriate menu selection LCD refreshment mode d REF alarm audio indication AUDI and clear the minimum and maximum memory CLR ATAL instruments passed the following electromagnetic compatibility EMC tests radiation EN 55022 class B immunity EN 61000 ...

Reviews: