Atal ATC-06 Instruction Manual Download Page 3

3

Battery voltage drop below 7 V is indicated with blinking of "BAT" in default display mode 

(displaying of actual values) and FAST mode is automatically canceled to save the battery. At the 
same time audio indication of alarms is automatically switched OFF.

If  instrument  is  powered  from  external  ac/dc  adapter,  internal  9V  battery  is  replaced  with 

rechargeable NiCd accumulator. In usual operation from adapter accumulator is charged only with 
small  current.  If  accumulator  is  totally  discharged,  its  full  charging  in  instrument  takes 
approximately  100  hours.  Instrument  with  accumulator  is  not  recommended  for  permanent 
operation without ac/dc adapter plugged. Accumulator works only as a standby source in case of 
power mains failure.

Switching ON and OFF the instrument

Switch ON the instrument by pressing ON/OFF key. After switching ON 
the instrument all symbols on the LCD are displayed. If the ON/OFF key is 
being held pressed, all LCD symbols are displayed till the key is released. 
In usual operation instrument then starts the measurement mode and actual 
measured values are displayed.

Displaying of actual measured values

In this mode is instrument anytime after switching ON. It is possible to enter this mode from 

other modes by pressing or by repeating pressing of MENU key.

Temperature  in  °C  is  displayed  on  the  upper  LCD  line  and  relative 
humidity in %RH is displayed on the lower LCD line.

Press DISPLAY key to display other readings - temperature in °C on upper 
line and dew point temperature on lower line (°C DP).

Function HOLD (storing of actual measured values) and minimum a maximum memory

Press  HOLD    key  in  the  default  mode  (displaying  of  actual  measured  value)  to  store  actual 

measured values to internal memory (indicated by short beep). Anytime it is possible to display 
stored  values  from  MENU  (see  below).  Each  pressing  of  the  HOLD  key  in  the  default  mode 
causes  values stored in HOLD memory are replaced with actual ones.

Switched ON instrument permanently updates minimum and maximum 
memory of each measured values. Press MIN key (resp. MAX key) in the 
default  mode  to  display  minimum  (resp.  maximum)  reading.  These 
minimum  and  maximum  readings  are  indicated  by  MIN  (MAX)  symbols 
on  the  LCD.  Press  DISPLAY  key  to  display  minimum  (resp.  maximum) 
value of other values. Pressing MIN (MAX) or MENU key again to return 
to  default  mode.  Minimum  and  maximum  memory  is  cleared  from  menu 
after confirmation selection CLR (see below). Values in HOLD, MIN and 
MAX memories remain stored even after instrument is switched OFF.

Summary of Contents for ATC-06

Page 1: ...ATAL ATC 06 Digital thermometer hygrometer Instruction manual ...

Page 2: ...ttery 9V or ac dc adapter 12V with NiCd accumulator 9V Average current consumption 0 15 to 0 7 mA depending on operation mode Typical life of Zinc Chloride battery 4 months Typical life of Alkali Mangan battery 6 months Operation conditions Ambient temperature range 10 to 60 C Ambient relative humidity range 5 to 95 RH Outer characteristics in accordance with EN 33 2000 3 normal environment with c...

Page 3: ...nt anytime after switching ON It is possible to enter this mode from other modes by pressing or by repeating pressing of MENU key Temperature in C is displayed on the upper LCD line and relative humidity in RH is displayed on the lower LCD line Press DISPLAY key to display other readings temperature in C on upper line and dew point temperature on lower line C DP Function HOLD storing of actual mea...

Page 4: ...NTER key to display values stored in the HOLD memory Press DISPLAY key to display other stored values dew point temperature Press MENU key to leave this mode Battery voltage of partially loaded battery is displayed This value illustrates battery condition Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode ...

Page 5: ... the arrow key UP to make value increase fast Press and hold the arrow key DOWN to make value decrease fast Release the arrow key and press ENTER to confirm new limit Reading ALHI indicates adjusted upper limit of the same alarm here air temperature Set the desired value in the same way as in above lower limit If needed it is possible to get back to lower limit setting of the same alarm by pressin...

Page 6: ...battery for longer than 1 minute even if instrument is switched OFF If it happens or if battery is totally discharged it is necessary to set again in appropriate menu selection LCD refreshment mode d REF alarm audio indication AUDI and clear the minimum and maximum memory CLR ATAL instruments passed the following electromagnetic compatibility EMC tests radiation EN 55022 class B immunity EN 61000 ...

Reviews: