ATAGO RX-7000a Standard Operating Procedure Download Page 4

 

 

   

P-SOP0804-3262-E02 

4/5 

Standard Operating 

Procedure    

SOP

 

 

 

Automatic digital 

Refractometer 

RX-7000

α

(alpha) 

 

 

 

6. In the event of a malfunction (failure) 

 

When operational malfunction (abnormalities) of the RX-

7000α occurs, 

check the operation of 

the RX-

7000α

 by the following procedure. 

After turning on the RX-

7000α,  c

heck the technical malfunction by "2. Daily check" and 

measurement of the solution of a known concentration. (example: Brix 10%) 

 

6-1. Check Procedure 

1) Turn off the power supply switch of the RX-

7000α    and turn on again

  2) Confirm the initial screen (opening message) and the present temperature in the   

      measurement display screen. 

3) Perform zero adjustment with distilled water. 

4) Measure the solution of a known concentration. (example: Brix 10%) 

In the above procedure, operational malfunction (abnormalities) can be checked or found. 

Properly identify the situations  of operational malfunction  (abnormalities) of the   

RX-

7000α

 correctly. 

 

 

6-2. Proper care 

Depending on the operational malfunction (abnormalities) of the RX-

7000α

, decide proper 

care to correct. 

The cause of operational malfunction (abnormalities) can be divided roughly into two areas. 

 

1)

 

The measured value is lower or higher due to an error in operation, mistake, etc   

    a.    Sample from the previous measurement may have been left on the prism surface. 

b.

 

Distilled water was left on the prism surface. 

c.

 

Measurement was executed without closing the cover lid. 

2)

 

If the operator of the RX-

7000α

  makes any mistakes in operation or inputting  of    

work, the RX-

7000α 

displays the corresponding error message. 

a.

 

Zero adjustment was performed with a sample. 

b.

 

Measurement was started with nothing on the prism, or zero adjustment was 

started nothing on the prism . 

In case of the above malfunctions, the end user can resolve by ensuring the above points are 

checked and avoided.  Carry out operation and measurement again.  Perform step 3 

(Periodical inspection of the preceding clause). 

 

3)  Failure in machinery or parts 

When the RX-

7000α does not operate

 properly due to mechanical trouble,   

under the same measuring conditions, degradation of the parts inside the unit and/or 

failure is assumed.    The RX-

7000α

 may be in need of repair. 

 

     a. Consult with the supervisor of the refractometer.   

     b. Record the trouble on an operating work report. 

     c. Contact the company where the refractometer was purchased. 

d. Contact the manufacturer of the refractometer.    (Make a contact list of the   

company where the refractometer was purchased along with the manufacturer.) 

     e. In the case of repair, it becomes impossible to use the refractometer.    Thus,       

  decide on methods of care and put them in writing. 

 

 

 

 

Summary of Contents for RX-7000a

Page 1: ...more distilled water onto the prism surface and close the cover lid then press the SW1 ZERO key Screen displays WAITING After the current temperature PRESENT reaches the targeted temperature TARGET the screen displays READING After a few seconds screen displays ZERO SET END Confirm the display of ZERO SET END Wipe off the distilled water from the prism surface with tissue paper 3 Drop 0 1ml or mor...

Page 2: ... point is confirmed by the calculation software of the refractometer that is all for calibrating all measuring spans 3 Periodical Inspection Inspect the RX 7000α with standard sucrose solution once a month to see if it indicates the measured value correctly Select a concentration point which you often measure and use this as a reference for inspection Save the checked data and record 3 1 Inspectio...

Page 3: ...s completed be sure that nothing on the prism surface and surrounding areas are completely wiped off with tissue paper moistened with water then wipe dry with tissue 2 When the body of the instrument is dirty wipe it with dry tissue cloth etc 3 If the digital printer is used the thermo sensitive roll paper is a consumable item 4 Place the refractometer at a designated location Keep it in a place o...

Page 4: ...ration mistake etc a Sample from the previous measurement may have been left on the prism surface b Distilled water was left on the prism surface c Measurement was executed without closing the cover lid 2 If the operator of the RX 7000α makes any mistakes in operation or inputting of work the RX 7000α displays the corresponding error message a Zero adjustment was performed with a sample b Measurem...

Page 5: ...ometer RX 7000α alpha 7 Establishment and record of revision When this standard operation procedure document is revised record the revised or changed contents in this sheet and sign approval column Record of revision Approval column Author s signature Approved by Date Establishment ...

Reviews: