background image

 

 

   

P-SOP0804-3262-E02 

3/5 

Standard Operating 

Procedure    

SOP

 

 

 

Automatic digital 

Refractometer 

RX-7000

α

(alpha) 

 

 

 

3-2. Inspection by nD (refractive index) 

The RX-

7000α

 has the thermostatic function.   

Measure distilled water by setting the targeted temperature to 10.0, 20.0, 30.0, 40.0, 

50.0, 60.0, and 70

  respectively. 

 

Be sure to perform zero adjustment with distilled water at each temperature.   Then, 

measure distilled water.    Confirm that the measured value coincides  with  "each 

temperature of the refractive index of distilled water". 

 

 

   

Accuracy :   

Under measurement temp. 10.0 to 30

 

     

±

0.0002 

  Under measurement temp. 31.0 to 70

   

     

±

0.0003        

   

 

 

 

 

 

 

4. Calibration 

 

  1)  Perform the calibration once a year. 

2)  Perform in accordance with the performance test method of operational qualification (OQ). 

3)  Record the data and other items in a performance test report, and have an operator 

and approver’s  signature.   The manufacturer  of  the  refractometer  may also be 

requested to perform calibration. 

 

    

 

 

 

 

 

 

 

 

 

 

 

 

5. Maintenance 

 

1)

 

When measurement is completed, be sure that nothing  on the prism surface and 

surrounding areas are completely wiped off with tissue paper moistened with  water, 

then wipe dry with tissue. 

2)

 

When the body of the instrument is dirty, wipe it with dry tissue, cloth, etc. 

3)

 

If the digital printer is used, the thermo sensitive roll paper is a consumable item. 

4)  Place the refractometer at a designated location. Keep it in a place of relative humidity 

less than 75%

, if it is not to be used for a long period of time. 

 

 

Temperature 

Refractive index 

10.0

 

1.33369 

20.0

 

1.33299 

30.0

 

1.33194 

40.0

 

1.33061 

50.0

 

1.32904 

60.0

 

1.32725 

70.0

 

1.32533 

When calibrating the RX-

7000α 

at your company (end user), it is required to meet the 

difficult conditions  (maintenance and check for environmental conditions, examination 

conditions and the accuracy of standard liquid, and the traceability of the jigs used) for 

calibration. 

In order to calibrate the RX-

7000α

, enormous expense and time may be spent. 

When the  matter  of expense  is taken into consideration, it may be more benefical to 

request the manufacturer of the refractometer to carry out calibration. 

* In that case, contact the company where the refractometer was purchased. 

 

 

 

 

 

Summary of Contents for RX-7000a

Page 1: ...more distilled water onto the prism surface and close the cover lid then press the SW1 ZERO key Screen displays WAITING After the current temperature PRESENT reaches the targeted temperature TARGET the screen displays READING After a few seconds screen displays ZERO SET END Confirm the display of ZERO SET END Wipe off the distilled water from the prism surface with tissue paper 3 Drop 0 1ml or mor...

Page 2: ... point is confirmed by the calculation software of the refractometer that is all for calibrating all measuring spans 3 Periodical Inspection Inspect the RX 7000α with standard sucrose solution once a month to see if it indicates the measured value correctly Select a concentration point which you often measure and use this as a reference for inspection Save the checked data and record 3 1 Inspectio...

Page 3: ...s completed be sure that nothing on the prism surface and surrounding areas are completely wiped off with tissue paper moistened with water then wipe dry with tissue 2 When the body of the instrument is dirty wipe it with dry tissue cloth etc 3 If the digital printer is used the thermo sensitive roll paper is a consumable item 4 Place the refractometer at a designated location Keep it in a place o...

Page 4: ...ration mistake etc a Sample from the previous measurement may have been left on the prism surface b Distilled water was left on the prism surface c Measurement was executed without closing the cover lid 2 If the operator of the RX 7000α makes any mistakes in operation or inputting of work the RX 7000α displays the corresponding error message a Zero adjustment was performed with a sample b Measurem...

Page 5: ...ometer RX 7000α alpha 7 Establishment and record of revision When this standard operation procedure document is revised record the revised or changed contents in this sheet and sign approval column Record of revision Approval column Author s signature Approved by Date Establishment ...

Reviews: