P-OQ1011-3276-J01
7
Operational
Qualification (OQ)
Automatic Digital
Refractometer RX-5000i
1. nD [1.33299] DATA [0.00]
2. nD [1.34782] DATA [10.00]
3. nD [1.36384] DATA [20.00]
4. nD [1.38115] DATA [30.00]
5. nD [1.42009] DATA [50.00]
4)
Touch “%mass” in “UNIT” and touch “ENTER.” Programming the user scale “Brix050”
is now complete. The scale-edit screen will appear, and the “Brix050” scale will be
displayed. Touch “ENTER.” The scale name list will be displayed. Touch “HOME” to
return to the home screen. Touch “MEASUREMENT” to display the measurement
screen.
5)
Touch “MODE” to display the MODE screen. Touch “MODE” and then “3.” Touch
“WAIT TIME” and then “5.” Touch “ENTER.” The measurement screen will appear.
Make sure that “MODE 3” and “WAIT TIME 5S” are displayed.
6)
Place a round sheet and apply the 1.51 nD standard liquid on the prism and touch
“START.” “WAITING TIME” will be displayed, and the countdown starts from “05.”
Once it reaches “00,” “READING” is displayed on the measurement screen. If “OUT
OF SCALE” is displayed in a few seconds, the RX-5000i is functioning properly. Wipe
the solution off the prism.
3-4. Printer Output Test (When Printer is Connected)
1)
Referring to the instructions in “1. Instrument Preparation” above, connect a printer
and turn the power on. Turn the power on to the RX-5000i. Make sure that the
opening message is printed.
2)
Apply distilled water on the prism and zero-set the instrument. Make sure that
“ZERO SET END” is printed.
3)
Measure the 30.00% Brix standard sucrose solution. Make sure that the sample
number, measurement values in nD and Brix, and measurement temperature are
displayed. (The items to be printed are selected in “1. Instrument Preparation.”)
3-5. RS-232C Output Test (When RS-232C Cable is Connected)
For Windows 95 / 98 / 2000 / XP
1) Connect the RX-5000i and a PC with a RS-232C cable, referring to the instruction
manual. Make sure that both units are turned off while connecting the cable.
2) From the PC menu, select Accessories > Communications > HyperTerminal. Connect
directly to COM1. Set up the communication parameters as follows:
●
Baud rate: 9600
●
Data: 7 bit
●
Stop: 1 bit
●
Parity: odd
3) Set the baud rate of the RX-5000i in the same manner. Touch “SET UP” on the home
screen. Touch “SERIAL,” and the protocol setting screen will appear.
4) Apply distilled water on the prism and zero-set the instrument.
5) Type in the zero-setting command “*ZERO” and press “ENTER” on the PC. Zero-setting
will be carried out.
6) Measure the 30.00% standard sucrose solution. The measurement data will be
displayed on the PC. Make sure that the format is as outlined in the instruction