P-OQ0612-5291-E03
7/7
Operational
Q u alification (O Q
)
Automatic Polarimeter
AP-300
Operation Qualification Results Report
Items
Descriptive contents
Remarks
Date of test
Product name
and model name
Serial number
Printer name
Serial number
Environmental
conditions
Temperature °C
Relative humidity %
1. Performance Test Data
Standard value
Standard
quartz
control plate (34.77°)
standard sucrose
solution of 100°Z
1st test
2nd test
3rd test
4th test
Average value
Difference
Judgment standard Accuracy Angle of rotation ±0.03 °
2. Function Examination
Functions
Contents
Switch function
□ Power supply switch
Key functions
□ZERO □QUIT □MENU □SCALE
□START □ENTER □Arrow keys
□NUMERIC keypad
Display functions
□ Opening message
□ Date and time
Error messages
□ ZERO SET END □OUT OF RANGE
□ TEMP. ERROR
Output functions
□ Printer output □RS
-232C output
*check each box
□.
3. Evaluation
The results of the performance test and a functional examination were evaluated.
□ Pass □ Fail
Date of operation: Year Month Day
Date of approval: Year Month Day
Operator signature: ___________________
Supervisor signature: ____________________