26
(2) Setting alarms
①
Choose the measurement scale for which the alarm
is to be set (nD, Brix, or CONC).
The example below is for setting the alarm in Brix
measurement mode.
②
Press the START/STOP key to pause the
measurement.
③
Press the SCALE key three times.
The SET UP light will then blink, and "SEt" will be
displayed on the measurement value display (Fig.
11-1).
④
Press the ENTER key and the SET UP light comes
on.
Using the
↑ ↓
keys, select "2. ALAr" (Fig. 11-7).
⑤
Press the ENTER key to enter the Alarm setting
screen.
Fig. 11-7
⑥
When L.OVER lights comes on.
Use the
↑ ↓
keys to change the value, and
select the decimal place with the
← →
keys (Fig.
11-8).
※
Press the SCALE key to return to the previous
"2. ALAr" screen.
⑦
Press the ENTER key to set the selected value.
The H.OVER setting will appear next.
Adjust the H.OVER value in the same manner as in
step
⑥
and press the ENTER key to confirm.
Fig. 11-8
⑧
Set the period of time the measurement values can
continue to exceed programmed upper and lower
limits, after which the alarm will be activated.
Use the arrow keys to select from 0 to 99 seconds.
The temperature display will then indicate "SEC"
(Fig. 11-9).
⑨
Press the ENTER key to confirm the changes.
The display will return to "2. ALAr ".
Fig. 11-9