24
11. Setting Measurement SET UP Mode
(1) Calibration Procedure
①
Make sure that the detection section is properly
mounted on the tank.
②
Let purified water or a reference sample flow into
the tank.
※
If the last measurement value was LLLL.LL or
HHHH.HH, the screen for setting the bias will not
appear.
③
Turn on the power as instructed in "9.Turning On
the Power," on page 20.
④
Perform calibration in the appropriate measurement
scale for your sample (i.e. nD, Brix, or Conc).
The example below is for calibration in Brix
measurement mode.
⑤
Press the START/STOP key to pause the
measurement.
⑥
Press the SCALE key three times.
The SET UP light will then blink, and "SEt" will be
displayed on the measurement value display (Fig.
11-1).
※
If left inactive for 10 seconds, the display returns to
the "StoP" screen, following three short error beeps
(Fig. 11-2).
Fig. 11-1
Fig. 11-2