(3) Setting recorder output
①
Press the START/STOP key to pause the
measurement.
②
Press the SCALE key three times.
The SET UP light will then blink, and "SEt" will be
displayed on the measurement value display
.
③
Press the ENTER key and the SET UP light comes
on.
Using the
↑ ↓
keys, select "3. rEC" (Fig. 11-10).
④
Press the ENTER key to enter the Recorder output
setting screen.
Fig. 11-10
⑤
The temperature display will indicate "Lo" (Fig.
11-11), and the screen for setting the recorder
output lower limit will appear.
Use the arrow keys to change the value.
※
Press the SCALE key to return to the previous
"3. rEC" screen.
Fig. 11-11
⑥
To confirm the changes, press ENTER.
The temperature display will indicate "Hi" (Fig.
11-12), and the screen for setting the recorder
output upper limit will appear.
Use the arrow keys to change the value.
※
Press the SCALE key to cancel the changes and
return to the "3. rEC" screen.
⑦
Press the ENTER key to confirm the changes.
The display will return to "3. rEC".
Fig. 11-12