![ATAGO 3534 Instruction Manual Download Page 17](http://html.mh-extra.com/html/atago/3534/3534_instruction-manual_2995846017.webp)
15
Mounting procedure
①
Install the instrument so that the prism
surface is at a right angle to the ground.
②
Attach the sample inlet unit to the instrument
with O-ring (accessory) inserted between
them, and fasten them together with the
clamp band (accessory).
③
Install the inlet unit so that the sample
solution runs from the lower nozzle to the
upper nozzle to prevent air bubbles from
forming.
④
When connecting the tubes to the hose
connector, clamp them with a tie band.
C
"Use of tie bands (Hose connector
(RE-67501) only) " page 16
⑤
The prism surface may become contaminated
with solids, dirt and/or grease. If this happens,
the prism surface must be cleaned by hand.
C
"16.Cleaning the Prism" page 32
The sample inlet unit should be installed in
such a manner that it can be easily removed
to allow access to the prism for cleaning.
N
Suspending the instrument (2.4kg) when
using the sample inlet unit to connect to
the piping is dangerous.
Use of the optional stand is
recommended.
C
"5.Mounting the instrument and AC
adapter AD-32 (AD-33 or AD-34)
(Optional)" page 13
C
"18.Consumable Parts and Optional
Items" page 34
Fig. 6-2
Fig. 6-3
Fig. 6-4
Top
Bottom Prism stage
Flow of
solution
O-ring
Flow of solution
Clamp band
Prism stage
O-ring
(silicon or EPDM)
Sample inlet unit
(optional)