P23
compromise the system's effectiveness, as well as the safety of the patient, you,
and others.
•
Only use this system if you have read, understood, and know all the safety
information, safety procedures, and emergency procedures contained in this
"Safety" section. Operating the system without proper awareness of safety use
could cause fatal or other serious personal injury.
Product Compatibility
Do not use your system in combination with other products or components,
unless expressly recognized by ASUSas compatible.
For information about such products and components, contact your
ASUSrepresentative.
Changes or additions to the system should be made only by either ASUSor third
parties expressly authorized by ASUSto do so. Such changes and additions must
comply with best engineering practice and all applicable laws and regulations with
the force of law within the jurisdictions concerned...
Equipment Protection
Follow these warnings to protect your system:
•
DO NOT immerse the probe into any liquid beyond the immersion level. In no
circumstances immerse the probe connector into any liquid.
•
DO NOT drop the probe or subject it to any type of physical shock or impact.
Impaired performance or damage such as housing cracks or chips could result.
•
Do NOT modify this device without the authorization of Leltek.
•
Do NOT use the probe with high frequency surgical equipment. Such use may
damage the device.
•
Do Not use the product in proximity to a strong electromagnetic field,
electromagnetic wave or in a magnetic environment. Such use may lead to a
possibility of measurement errors or product damage.
•
While the LU700 Series device is being charged using a mobile charging power
supply, do NOT use for diagnostic work.
•
The LU700
Series device should be charged using the mobile charging power
supply and medical product charging cable in compliance with IEC 60601-1 for
two MOPP insulation system. The mobile charging power supply should be
checked and replaced regularly.
If the system or transducers have been in an environment in excess of 35°C (95°F),
allow them to cool to operating temperature before initiating the system or
connecting the transducers.
Do Not allow the transducer to contact the patient if the temperature of the
transducer is higher than 43°C (109°F), the transducer will automatically shut
down. Please let the transducer to cool the extent possible. If the transducers