850A/B User Manual Publication No. 980972 Rev. B
Contact Current Adjustment
1. Connect a current meter to Channel 1 Input. Set to read up to 11 milliamps.
2. Place the X1-X0.1 switch to X1.
3. Place the CONTACT CURRENT DIAL to 10.
4. Set the full-scale current adjust R32 for 10 ±.01 milliamps indication on the
current meter.
5. Connect the current meter to each of the other nine channels and get 10 ±0.2
milliamps indication.
6. Place the X1-X0.1 Switch to X0.1.
7. All channels should indicate 1.0 ±0. 1 milliamps.
8. Set dial to 5, and X1-X0.1 Switch to X1.
9. All channels should indicate 5.0 ±0.1 milliamps.