Publication No. 980956 Rev. A 346B User Manual
Astronics Test Systems
Performance Test Procedure 3-3
20. Connect the calibrator mV OUTPUT to Distortion Analyzer to verify Distortion is
equal or less than the listed values with the mV OUTPUT level set at 9999
millivolts rms.
20 Hz
.25% (max)
100 Hz
.25% (max)
1 kHz
.25% (max)
10 kHz
.25% (max)
50 kHz
.5% (max)
100 kHz
1.0% (max)
File: 99-Performance Test Program
SEGMENT FREQUENCY
MODE
LEVEL
UNITS
TIME
NOTES
01
00346.2
SIN
XXX.0
RMS
5
IS SEG
XXX IS THE
SERTIAL NO.
2
-
DC
9999
DC
5
SEG 2-5: SEG
3
-
DC
-999.9
DC
5
LEVEL
CHECKS
4
-
DC
99.99
DC
5
5
-
DC
9.999
DC
5
6
00100.0
SIN
9999
RMS
5
FREQ
RESPONSE
CHECKS
7
99999.9
SIN
9999
RMS
5
8
10,000.0
SIN
9999
RMS
5
9
1000.0
SIN
9999
PEAK
5
MODE
UNITS
CHECKS
10
1000.0
SIN
9999
PK-PK
5
11
1000.0
SQ
9999
PK-PK
(AVG)
5
12
1000.0
TRI
3464
PK-PK
(AVG)
5
13
00346.2
SIN
3464
RMS
0