Commands and Parameter
LDM41E and LDM42E Manual
Page 32
ASTECH GmbH
8.18
HF – display/set Temperature for heating off
15
HF displays/sets the temperature value at which the internal heating is
switched off. The values can be varied between -40°C and 70°C. Standard
setting is 12°C.
8.19
RMx y z – display/set remove measurement
Standard setting: 0 0 0
RM is intended to facilitate settings for a range of expected distance values.
Values which are found to be outside of this expected range will be corrected
until matching the most recently valid measured values.
RM is only effective in DT mode.
It consists of three parameters which are separated by space (20h).
x
designates the number of preceding measured values that will be
evaluated in the case of non-conforming measurement. A maximum
of ten preceding measured values can be evaluated.
y
defines the range of permissible values. If this range is exceeded in
negative or positive direction, the respective measured value will be
corrected accordingly.
z
stands for the number of values that are out of the permissible value
range (out of tolerance values). In the event of out-of-tolerance
values arriving in succession, the most recently corrected value will
be included in the correction process for the next out-of-tolerance
value. The maximum allowed number of out-of-tolerance values is
100.
Important: The use of RM parameter settings should be restricted
to suitable applications only. Improper use of the parameter may
create safety hazards!