3. 

Specifications 

 

3.1.  Technical data 

Number of measurements in memory 

storage 

500 

Clock rate deviation  < 1 minute/month 

Reading range  up to 20 mm 

Power supply  - integral lithium polymer rechargeable battery 3.7 V, 2Ah 

- USB connector 

Battery life  - sufficient for 2,500 measurements 

Max. current consumption  - at USB port 105 mA 

- on battery charger 500 mA 

Max. charging time  - at USB port only partial charging within 6 hours 

- on battery charger 6 hours 

Ambient conditions  -20° C to + 60° C 

 
 

3.2.  Dimensions/ connections 

Housing material: Polyamide 
Dimensions: 

  Width: 115 mm 

  Length: 206 mm (display 85 mm) 

  Height: 85 mm 

Connection: Micro USB 2.0 Type B (socket) 
 

4. 

Operation oft the appliance ADL 280 RFID 

 

4.1.  Power supply 

Power  is  supplied  to  the  Data  Logger  by  means  of  an  integral  rechargeable  battery.  If 
connected up to the USB port of a PC or a USB charger, power is supplied via the USB. This 
ensures  immediate  operating  availability  even  if  the  battery  is  flat  while  also  facilitating 
charging of the integral battery. 
The  charging  level  of  the  integral  battery  can  be  gauged  via  the  symbol  appearing  on  the 
bottom left-hand side of the display. As soon as the battery is almost flat, an alert message will 
be displayed. 
 

A.S.T. GmbH
ADL 280 RFID

-- Batterie leer --

28.01.15 17:32:58

 

Figure 1

 – 

Alert message on start screen indicating that the battery is flat

 

 

Any further discharging of the battery will result in the appliance automatically switching itself 
off although the integral clock will continue to function for the time being. 
 

Summary of Contents for ADL 280 RFID

Page 1: ... T Angewandte System Technik GmbH Mess und Regeltechnik Data Logger ADL 280 RFID Original Operating Manual A S T Angewandte System Technik GmbH Mess und Regeltechnik Marschnerstraße 26 01307 Dresden Telefon 03 51 44 55 30 Telefax 03 51 44 55 555 www ast de ...

Page 2: ...Dimensions connections 4 4 Operation oft the appliance ADL 280 RFID 4 4 1 Power supply 4 4 2 Operating controls 5 4 3 Implementation of measurement procedure 7 4 4 Display of measurement results 7 4 5 Switching off the appliance 7 5 Operation of software xks280 7 5 1 Installation of software xks280 7 5 2 Reading out data from Data Logger ADL 280 RFID 7 5 3 Configuration of appliance 9 5 4 Addition...

Page 3: ...e manufacturer 2 Method of functioning The Data Logger ADL 280 RFID is a reading device designed for use in conjunction with an A S T RFID transponder Power is supplied to the transponder by means of an alternating magnetic field with a frequency of 125 kHz The transmission of commands to the transponder takes place via amplitude modulation of the alternating magnetic field response being accompli...

Page 4: ...t 85 mm Connection Micro USB 2 0 Type B socket 4 Operation oft the appliance ADL 280 RFID 4 1 Power supply Power is supplied to the Data Logger by means of an integral rechargeable battery If connected up to the USB port of a PC or a USB charger power is supplied via the USB This ensures immediate operating availability even if the battery is flat while also facilitating charging of the integral b...

Page 5: ...Operating controls and display functions on the Data Logger The pushbutton controls trigger the following functions Pushbutton Action Function Press briefly 5 sec Previous dataset Press and hold 5 sec First dataset beginning of list Press briefly 5 sec Switches on the appliance Performs measurement pressing a second time Press and hold 5 sec Switches off the appliance Press briefly 5 sec Next data...

Page 6: ...ul the display will contain the following information Figure 4 Anzeige eines Messdatensatzes A S T GmbH ADL 280 RFID frei 495 500 21 01 15 17 32 58 Manufacturer Designation of appliance Enough storage space remaining for dataset Clock time format Hours minutes seconds Date format Day month year Battery display with symbols Total storage space available Power supply via USB Battery topped up Batter...

Page 7: ...asurement result Figure 4 4 5 Switching off the appliance The appliance is switched off by pressing and holding down the key for longer than 5 seconds or automatically after approximately 2 minutes inactivity 5 Operation of software xks280 5 1 Installation of software xks280 The software package xks280 calls for Microsoft NET Framework 4 Client Profile If this is available on the PC the software c...

Page 8: ...ut takes place automatically Figure 6 Software xks280 Display of list showing all datasets The internal list can be sorted according to the content of the column by clicking onto the column header The columns are modifiable according to their order by clicking onto them with the mouse and dragging them as required 3 Saving the data a Export to Excel Save in Excel program b as comma separated value...

Page 9: ...ompletely emptied from the Data Logger ADL 280 RFID by actuating the command button Delete from appliance CAUTION The data should first be saved before pressing the command button Delete from appliance Data deleted from the ADL 280 RFID are irretrievably lost 5 3 Configuration of appliance The appliance is configured by performing the following steps 1 Starting up of software xks280 2 Connection o...

Page 10: ...ntry of the date and time and activating the command Write data in device Note After the date and time has been successfully altered on the appliance a corresponding message will appear briefly within the status bar of the program window 4 Switchover to Tools window Figure 9 Software xks280 Calling up the tools 5 Selection of language on the Data Logger Figure 10 Software xks280 Selection of langu...

Page 11: ...s are as follows The Data Logger must be connected up to the PC The transponder must be located within the reading range of the Data Logger The field on the Data Logger must be switched on RF Field button Due to limited power consumption at the USB port charging of the integral battery is interrupted whenever the magnetic field is switched on Figure 11 Software xks280 Tools The following functions...

Page 12: ...Shielding of magnetic field Remove metallic objects away from the magnetic field Interference affecting magnetic field Eliminate source of interference e g additional RFID reader or move away from surrounding area Insufficient magnetic coupling Vary the distance between the sensor and Data Logger or alter the position of the Data Logger Appliance not detected by the software Contact problems Check...

Reviews: