![Associated Research SC6540 Operation And Service Manual Download Page 45](http://html1.mh-extra.com/html/associated-research/sc6540/sc6540_operation-and-service-manual_2989485045.webp)
41
If a SC6540 main is configured with either two high current modules or two
high voltage modules, module A will be assigned a 0 address for channels
1-8 and module B will be assigned a 1 address for channels 9-16.
A system requiring 80 test points would require 4 secondary scanners each set up
with two 8 channel modules. Please refer to the following chart for circuit address
set up instructions:
SCANNER
TYPE
POSITION
CIRCUIT ADDRESS
CHANNELS
Main
A
Automatically assigned 0
1-8
Main
B
Automatically assigned 1
9-16
Secondary 1
A
2
17-24
Secondary 1
B
3
25-32
Secondary 2
A
4
33-40
Secondary 2
B
5
41-48
Secondary 3
A
6
49-56
Secondary 3
B
7
57-64
Secondary 4
A
8
65-72
Secondary 4
B
9
73-80
For the above configuration the DIP switches would be set as follows:
Main The DIP switch would be set for the GPIB Address
Secondary 1
Position A
PIN 1 = OFF
Address 2
PIN 2 = ON
PIN 3 = OFF
PIN 4 = OFF
Position B
PIN 5 = ON
Address 3
PIN 6 = ON
PIN 7 = OFF
PIN 8 = OFF
Secondary 2
Position A
PIN 1 = OFF
Address 4
PIN 2 = OFF
PIN 3 = ON
PIN 4 = OFF
Position B
PIN 5 = ON
Address 5
PIN 6 = OFF
PIN 7 = ON
PIN 8 = OFF
Secondary 3
Position A
PIN 1 = OFF