Associated Research HYPOT III 3705 Quick Start Manual Download Page 7

7

WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.

CONDUCT A TEST

TEST RESULTS

  

Connect the Interlock Disable Key (P/N 38075) to the signal input 

connector on the rear panel of the instrument. If you’re not utilizing a 

DUT enclosure (P/N 39064 or 39656) or other safety device, the Interlock 

Disable Key is required in order to run a test. 

  

With the instrument set to the desired test type and your DUT correctly 

connected to the instrument, you are now ready to start testing.

  

If the Continuity function is ON and the resistance of the ground circuit 

is less than the Max Resistance setting, the green TEST button will 

illuminate.

  

Push the green TEST button on the front panel. The DUT is tested for  

a duration equal to the Ramp and Dwell settings.

PASS:

  If the DUT passes the test, you will hear a short audible beep and the 

display will indicate the test results.

FAIL:

   If a failure occurs, you will hear a long audible alarm and the red flashing 

indicator will light up. Press the red RESET button to stop alarm.

Pass/Fail Indication Screen

HypotIII-QSG-MARCH17.indd   7

4/13/17   11:09 AM

Summary of Contents for HYPOT III 3705

Page 1: ... clips directly and always connect the return lead first You should always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN H...

Page 2: ... the rear panel of the instrument This is required to run a test Turn the instrument power switch ON The initialization screen will appear After three seconds the Perform Test screen will appear as shown below 3705 3765 3770 3705 3765 3770 3705 3765 3770 Perform Test Screen WARNING LOCATE A SUITABLE TESTING AREA WITH A THREE PRONG GROUNDED OUTLET BE SURE THAT YOUR THREE PRONG OUTLET HAS BEEN TESTE...

Page 3: ...ote OFF Single Step ON Alarm 5 Contrast 1 Results Last Lock OFF Mem Lock ON More Exit Select a parameter to edit by scrolling with the and soft keys Select the soft key to change the parameter M0 1 ACW Settings ____________ O 0OkV 10 OOuA O O00mA 1 Os Menu Memory Step Exit M0 1 ACW Settings ____________ O 0OkV 10 OOuA O O00mA 1 Os Menu Memory Step Exit START From the Perform Test screen select the...

Page 4: ...More Edit Exit Use the and arrow keys to select another parameter to edit Test Type ACW Voltage 1 24kV Max Lmt 1O OOmA Min Lmt O OOOmA Ramp UP O 1s Dwell 1 Os Ramp DN O Os More Edit Exit Select EDIT and then use the and soft keys to edit additional parameters Select ENTER to save changes for each parameter Test Type ACW Voltage 1 24kV Max Lmt 1O OOmA Min Lmt O OOOmA Ramp UP O 1s Dwell 1 Os Ramp DN...

Page 5: ...l Connect the black ground return lead P N 2100A 13 to the front panel return terminal and connect the other end of the lead to the dead metal on the chassis of the DUT Check to ensure a solid connection is made between the DUT and the return clip Plug the line cord of the DUT into the adapter box receptacle 5 WARNING DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST Note Be sure to enabl...

Page 6: ...ont panel of the instrument Connect the red high voltage lead P N 4040A 08 to the H V terminal Connect the clip end of the high voltage lead to the current carrying conductors of the DUT s circuity 6 TEST CONNECTION DUT Connections Exposed Metal Chassis Setup WARNING DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST If your chassis does not have any exposed metal you can wrap the enclosur...

Page 7: ...ly connected to the instrument you are now ready to start testing If the Continuity function is ON and the resistance of the ground circuit is less than the Max Resistance setting the green TEST button will illuminate Push the green TEST button on the front panel The DUT is tested for a duration equal to the Ramp and Dwell settings PASS If the DUT passes the test you will hear a short audible beep...

Page 8: ... key features of the Hypot III please consult the full Operation and Service Manual or call us toll free 1 800 858 TEST 8378 or 1 847 367 4077 2017 Associated Research arisafety com 3 2017 Safety Is Our Only Focus SCAN FOR QUICK START VIDEO FOLLOW US ...

Reviews: