EDIT SYSTEM PARAMETERS

PERFORM A GROUND CONTINUITY CHECK

Configure the instrument system parameters to your preferences. The instrument 

system parameters are global and will affect all tests that you perform regardless 

of memory location and memory step.

You can easily perform a Ground Continuity check on Class I products through the 

use of the adapter box (P/N 36544) provided with the instrument and ground return 

lead. To enable this function, you will need to set the instrument to run either an AC 

or DC Hipot test.

Note:

 If you need to preform a Ground Continuity check on your DUT, be sure to enable 

the Ground Continuity test before moving forward. 

3

From the Perform Test screen, 
select the MENU soft key.

DCW   Settings
M1-1     0.0s

______________

Menu

kV

uA

Select EXIT twice to save changes
and return to the Perform Test screen.

PLC Remote    OFF
Single Step  ON
Alarm           5
Contrast      1
Results       Last
Lock            OFF
Mem Lock      ON

     <More>

 

 

+

Exit

>

>

Select the SYSTEM soft key.

M1-1   DCW
  Settings

____________
    O.3OkV
    5OOOuA
     O.OuA
     O.Os

Results

Test

System

Exit

Select the MENU soft key.

PLC Remote    OFF
Single Step  ON
Alarm           5
Contrast      1
Results       Last
Lock            OFF
Mem Lock      ON

     <More>

 

 

+

Exit

>

>

Select a parameter to edit by 
scrolling with the       and      soft keys. 
Select the + soft key to change 
the parameter.

M0-1   ACW
  Settings

____________
    O.0OkV
  10.OOuA
  O.O00mA
     1.Os

Menu

Memory

Step

Exit

M0-1   ACW
  Settings

____________
    O.0OkV
  10.OOuA
  O.O00mA
     1.Os

Menu

Memory

Step

Exit

START

From the Perform Test screen, 
select the MENU soft key.

Select the TEST soft key.

Select Edit soft key. Use the + and - 
soft keys to select ON. 

Select ENTER to save changes. 
Select EXIT twice to return to the 
Perform Test Screen. 

Arc Sense     O
Continuit

y

   ON

Max Lmt    1.5OΩ
Min Lmt    O.OOΩ
Offset     O.50Ω
Connect      OFF

+

 -

Edit

Esc

Arc Sense     O
Continuit

y

   ON

Max Lmt    1.5OΩ
Min Lmt    O.OOΩ
Offset     O.50Ω
Connect      OFF

+

 -

Enter

Exit

DCW   Settin

g

s

M1-1     0.0s

____________________________

Menu

kV

uA

M1-1   DCW
  Settin

g

s

________________
    O.3OkV
    5OOOuA
     O.OuA
     O.Os

Results

Test

S

y

stem

Exit

Use the        arrow key to select 
CONTINUITY on the next screen.

Test T

y

pe    DCW

Volta

g

e     O.3OkV

Max Lmt     5OOOuA
Min Lmt      O.OuA
Ram

p

 UP      O.1s

Dwell        O.Os
Ram

p

 DN      O.Os

     <More>

 

 

Edit

Exit

>

>

Select the MENU soft key.

M0-1   ACW
  Settings

____________
    O.0OkV
  10.OOuA
  O.O00mA
     1.Os

Menu

Memory

Step

Exit

M0-1   ACW
  Settings

____________
    O.0OkV
  10.OOuA
  O.O00mA
     1.Os

Menu

Memory

Step

Exit

HypotIII-QSG-MARCH17.indd   3

4/13/17   11:09 AM

Summary of Contents for HYPOT III 3705

Page 1: ... clips directly and always connect the return lead first You should always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN H...

Page 2: ... the rear panel of the instrument This is required to run a test Turn the instrument power switch ON The initialization screen will appear After three seconds the Perform Test screen will appear as shown below 3705 3765 3770 3705 3765 3770 3705 3765 3770 Perform Test Screen WARNING LOCATE A SUITABLE TESTING AREA WITH A THREE PRONG GROUNDED OUTLET BE SURE THAT YOUR THREE PRONG OUTLET HAS BEEN TESTE...

Page 3: ...ote OFF Single Step ON Alarm 5 Contrast 1 Results Last Lock OFF Mem Lock ON More Exit Select a parameter to edit by scrolling with the and soft keys Select the soft key to change the parameter M0 1 ACW Settings ____________ O 0OkV 10 OOuA O O00mA 1 Os Menu Memory Step Exit M0 1 ACW Settings ____________ O 0OkV 10 OOuA O O00mA 1 Os Menu Memory Step Exit START From the Perform Test screen select the...

Page 4: ...More Edit Exit Use the and arrow keys to select another parameter to edit Test Type ACW Voltage 1 24kV Max Lmt 1O OOmA Min Lmt O OOOmA Ramp UP O 1s Dwell 1 Os Ramp DN O Os More Edit Exit Select EDIT and then use the and soft keys to edit additional parameters Select ENTER to save changes for each parameter Test Type ACW Voltage 1 24kV Max Lmt 1O OOmA Min Lmt O OOOmA Ramp UP O 1s Dwell 1 Os Ramp DN...

Page 5: ...l Connect the black ground return lead P N 2100A 13 to the front panel return terminal and connect the other end of the lead to the dead metal on the chassis of the DUT Check to ensure a solid connection is made between the DUT and the return clip Plug the line cord of the DUT into the adapter box receptacle 5 WARNING DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST Note Be sure to enabl...

Page 6: ...ont panel of the instrument Connect the red high voltage lead P N 4040A 08 to the H V terminal Connect the clip end of the high voltage lead to the current carrying conductors of the DUT s circuity 6 TEST CONNECTION DUT Connections Exposed Metal Chassis Setup WARNING DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST If your chassis does not have any exposed metal you can wrap the enclosur...

Page 7: ...ly connected to the instrument you are now ready to start testing If the Continuity function is ON and the resistance of the ground circuit is less than the Max Resistance setting the green TEST button will illuminate Push the green TEST button on the front panel The DUT is tested for a duration equal to the Ramp and Dwell settings PASS If the DUT passes the test you will hear a short audible beep...

Page 8: ... key features of the Hypot III please consult the full Operation and Service Manual or call us toll free 1 800 858 TEST 8378 or 1 847 367 4077 2017 Associated Research arisafety com 3 2017 Safety Is Our Only Focus SCAN FOR QUICK START VIDEO FOLLOW US ...

Reviews: