To prolong the working life of the bed and for trouble free operation it is recommended that routine inspections
should be performed by a person who is suitably qualified and well acquainted with the design, use and care of
ComfiMotion beds.
If inspection reveals that the safety of the bed is jeopardised in any way, it should be
IMMEDIATELY WITHDRAWN
FROM SERVICE, tagged as such and the owner notified, until repair or replacement has occurred.
INSPECTION:
•
A visual and functional inspection should be performed monthly
• Visually inspect electrical components: can you see damage to the hand control or power cord?
• Visually inspect mechanical components: can you see damage, chips or deformation to the frame, headboard
or accessories?
•
Is the hand control functioning as it should?
•
Are the castors, brakes and accessories functioning as they should?
PREVENTATIVE MAINTENANCE:
The following preventative maintenance procedures should be carried out every
12 months or as otherwise specified:
NOTE: Other than general maintenance activities all other service or repair work must be performed by qualified
and trained persons or agents approved by Aidacare.
•
Overall functional check that all hand control functions are operating the bed as expected.
•
Overall inspection for visible signs of wear, tear or damage to the integrity of electronics and Cables.
•
Actuators and all moving components should not make unusual noises.
NOTE: If you are unsure if your ComfiMotion bed is working correctly, Contact Aidacare or your local authorised
Aspire Dealer.
10. SERVICE & MAINTENANCE
Effect
Possible Cause
Verification
Corrective Action
BED DOES NOT
STAY IN PLACE
Locking castors
not preventing bed
movement.
Brake System is
malfunctioning.
Not all locking castors have
been engaged.
Brake pedal is in lowest
position, but bed does not
stay in place.
Engage all 4 x locking castors on each
corner of the bed.
If this does not work, contact your sales
representative or a qualified Aidacare
service technician.
BED DOES NOT
MOVE UP OR
DOWN
Bed is not plugged
in and optional
battery is flat.
Actuator is faulty.
Handset is faulty.
Check if the power lead is
plugged into mains power.
Swap actuators.
Swap handsets.
If unplugged, plug power leads into
mains power. Allow 3 hours to charge
battery (if fitted).
Contact your sales representative or a
qualified Aidacare service technician.
BACKREST
DOES NOT
MOVE UP OR
DOWN
Actuator code is
unplugged.
Actuator is faulty.
Handset is faulty.
Check if cables have come
loose from control box (under
top frame at head end).
Swap actuator.
Swap handsets.
If loose, fully plug cables into
control box.
Contact your sales representative or a
qualified Aidacare service technician.
11. TROUBLESHOOTING