8. CLEANING 

To ensure all infection control protocols are strictly adhered to, it is extremely important to follow the below 

steps to keep the product clean at all times.

Pump

• Wipe with a damp cloth that has been pre-soaked with a mild detergent.

• Ensure pump is kept free of dust and other particles.

 

 

DO NOT!

 Immerse or soak the pump at any time.

Mattress Cover

       • To remove small spills or stains, wipe the mattress cover with a damp cloth that has been  

          pre-soaked with a mild detergent or single use disinfectant wipe. 

       • To remove large spills or stains, the top cover should be completely removed and laundered as    

          per below procedure:

      1.  Remove top cover by unzipping all 4 x sides from the mattress base.

      2. Prewash top cover at up to 95

°

 for 15 minutes.

      3. Main wash top cover at up to 95

°

 for 15 minutes.

      4. Dry mattress cover prior to use by hanging up on a line/air dryer or on a low heat setting it can    

          be tumble dried for up to 90 minutes.

      5. Once dry, reattach top cover by zipping all 4 sides back onto the mattress base.

 

DO NOT!

 Use phenol based products.

 WARNING!

 After cleaning, the mattress must be completely dried prior to use.

 WARNING! 

Do not dry mattress cover in direct sunlight to prevent wearing out of text on cover.

WARNING

WARNING

9. MAINTENANCE & STORAGE

To ensure longevity of your Alternating Mattress Replacement, it is very important that it is correctly stored 

and regularly maintained. 

General

• Check mains power cable and plug for any abrasions or excessive wearing.

• Check mattress cover for signs of wear or damage.

• Check air hoses for any kinks or breakages .

Fuse Replacement

• Disconnect the mains power if there is a blown fuse.

• Remove the cover on the side casing of the pump.

• Remove old fuse and reinsert a new fuse, fix in place with the cover.

 

WARNING! 

Only use fuse supplied by your Aspire Dealer, ensuring it has the correct rating and  

 specification.

Air Filter Replacement

• Remove the cover on the rear casing of the pump.

• Remove old filter and reinsert a new filter, fix in place with the cover.

NOTE:

 If the filter does not require replacement, it can be cleaned by being washed gently with a mild 

detergent and warm water. Ensure the filter is dry prior to installing back into the pump.

 WARNING!

 Only use filter supplied by your Aspire Dealer, ensuring it has the correct rating and  

 

 specification.

Storage

• Release all air from mattress by 

OPENING CPR

 valve and disconnecting the mattress hose from the  

   pump.

• Lay the mattress out flat & roll from 

HEAD

 end towards 

FOOT

 end.

• Tighten mattress straps into clamps to prevent mattress unrolling.

• Packing strap can then be stretched around the rolled mattress to prevent unrolling.

WARNING

WARNING

Summary of Contents for ACTIVE AIR 8

Page 1: ...USER MANUAL 8 ALTERNATING MATTRESS REPLACEMENTS INCLUDE Single PTM639400T King Single PTM639410T...

Page 2: ...manual carefully before using your new mattress replacement TABLE OF CONTENTS 1 Introduction 2 Safety Precautions 3 Product Safeguards 4 Product Overview 5 Assembly Setup 6 Operation 7 Product Specifi...

Page 3: ...be used under the instruction of a carer or health professional where concerns exist around patient safe use Intended Use This product is intended to help and reduce the incidence of pressure injuries...

Page 4: ...nd size appropriate pressure output is automatically set ensuring optimal care and protection 24 Hour Transport In the event of a power failure patients can remain on the inflated surface for up to 24...

Page 5: ...s by placing on a soft surface such as a bed or couch Keep all valves tubes openings responsible for air flow clear of all lint hair and other particles 10 Do not insert any object into the air valves...

Page 6: ...es 5 ASSEMBLY SETUP 2 1 Box Contains 1 Digital Mattress Pump 2 Mains Power Cable 3 Mattress Replacement 4 Transport Carry Bag 5 User Manual When opening the carton ensure it is done in a safe and care...

Page 7: ...s switched OFF 10 Press down in the L shape power cord protector to secure the power cord in place 11 Plug the mains power cable into the electrical power socket 12 Set mode suitable for patient Refer...

Page 8: ...patient ingress egress and meal time It can also be used as a tool to completely inflate the mattress 1 Press the AUTO FIRM button down LED light will start flashing and the mattress will start to fir...

Page 9: ...button until the STATIC LED light is illuminated 2 This function will only work once Auto Firm is off 3 Static mode will automatically fall back to ALTERNATING mode after 20 minutes 1 2 1 2 2 1 NOTE T...

Page 10: ...er Function 1 Auto protect activates when the bed backrest angle exceeds 25 at this point an additional 10mmHg of pressure is provided by the pump This additional pressure increases to 15mmHg once bed...

Page 11: ...D light will illuminate and alarm 2 Refer to the Troubleshooting sticker on the base of the pump to identify potential solutions If the issue persists contact your Aspire dealer or Aidacare Customer S...

Page 12: ...6kg Cycle Times 10 15 20 25 30 minutes Static Time 20 minutes Auto Firm Time 20 minutes Pump unit Output Pressure Range mmHg 15 50 5 Voltage AC230 V 50Hz Current 0 23 AMAX 253V Fuse Rating T1AL 250VA...

Page 13: ...ARNING WARNING 9 MAINTENANCE STORAGE To ensure longevity of your Alternating Mattress Replacement it is very important that it is correctly stored and regularly maintained General Check mains power ca...

Page 14: ...Y SERVICE Aspire Active Air Acute 8 Alternating Mattress Replacements carry a comprehensive warranty as detailed below Please keep proof of purchase for proof of warranty commencement Consumer Warrant...

Page 15: ...tion Report Action taken Inspected By Date Date Date Date Date Signature Signature Signature Signature Signature YEAR 6 Service Type Condition Report Action taken Inspected By YEAR 7 Service Type Cond...

Page 16: ...30 31 NOTES...

Page 17: ...ASUM40 24 100321 Manufactured and Imported by AIDACARE PTY LTD Building 3A 1 Moorebank Avenue Moorebank NSW 2170 Australia 1300 133 120 www aidacare com au Made in Taiwan...

Reviews: