background image

 

Page 4 of 11 

2  Safety 

2.1  General sources of hazards 

Pressure  gauges are pressurized instruments  where failure can result in hazardous situations. The  selection  of 
pressure  gauge  should be made in accordance with the rules set out in EN 837-2

A failure resulting in injury or 

damage may be caused by excessive overpressure, excessive vibration or pressure pulsation, excessive instrument 
temperature, corrosion of the pressure containing parts, or other misuse. Consult the manufacturer before installing if 
there are any questions or concerns. 

2.1.1  O

verpressure

 

Pressure spikes

  in excess of the rated overpressure capability of the transducer may cause irreversible electrical 

and/or mechanical damage to the pressure measuring and containing elements. 

Fluid hammer 

and surges can destroy any pressure transducer and must always be avoided. A pressure snubber 

should be installed to eliminate the damaging hammer effects. Fluid hammer occurs when a liquid flow is sud

denly 

stopped, as with quick closing solenoid valves. Surges occur when flow is suddenly begun, as when a pump is turned 
on at full power or a valve is quickly opened. 

Liquid surges

  are particularly damaging to pressure transducers if the pipe is originally empty. To avoid damaging 

surges, fluid lines should remain full (if possible), pumps should be brought up to power slowly, and valves opened 
slowly. To avoid damage from both fluid hammer and surges, a surge chamber should be installed. 

Symptoms of fluid hammer and surge's damaging effects: 

• 

Pressure transducer exhibits an output at zero pressure (large zero offset). 

• 

Pressure transducer output remains constant regardless of pressure 

• 

In severe cases, there will be no output. 

2.1.2  Freezing 

Prohibit freezing of media in pressure port. Unit should be drained (mount in vertical position to prevent possible 
overpressure damage from frozen media. 

2.1.3  Static electrical charges 

Any electrical device may be susceptible  to damage when exposed to static  electrical charges. To avoid damage to 
the transducer observe the following: 

• 

Operator/installer should follow the  proper ESD (electrostatic discharge)  protection procedures before 

handling the pressure transducer components. 

• 

Ground the body of the device BEFORE making any electrical connections 

• 

When disconnecting, remove the ground LAST! 

 The shield and drain wire in the  cable (if supplied) is not connected to  the transducer body, and is not a 
suitable ground. 

 

2.2  Use in accordance with intended purpose 

The devices are only to be used for the intended purpose as described by the manufacturer. 

The devices are used for direct display of overpressures, vacuum  and  compound pressure  and  for transmission of 
overpressures, vacuum and compound pressure into a standard electrical output signal. 

For each use scenario, the corresponding set-up regulations must be respected.  

The usage in explosion risk areas is not allowed. 

2.3  Operator’s responsibility 

Safety instructions for proper operation of the device must be respected. They are to be provided by the operator for 
use by the respective personnel for installation, servicing, inspection and operation. Risks from electrical energy and 
from the released energy of the medium, from escaping media and from improper connection of the device must be 
eliminated. The details for this are to be found in the corresponding applicable set of regulations, such as DIN EN, 
UVV (accident prevention regulations) and in sector-specific instances of use (DVWG, Ex-. GL,  etc.) the VDE 
guidelines and the regulations supplied by local utilities companies. 

Summary of Contents for t5500e

Page 1: ...5 01EN Installation and Maintenance Instruction Manual Bourdon tube pressure gauge with integrated transmitter model T5500E for industrial application In the following configuration 100 T5500E bourdon...

Page 2: ...g on the device 5 5 Construction and function 6 5 1 Overview 6 5 2 Description of function 6 5 3 Description of components 6 5 4 Accessories 6 6 Transport 7 6 1 Safety 7 6 2 Transport inspection 7 6 3...

Page 3: ...ing This indicates a possibly hazardous situation where failing to follow advice may result in risks to people animals the environment and buildings Information This emphasizes key information for eff...

Page 4: ...e will be no output 2 1 2 Freezing Prohibit freezing of media in pressure port Unit should be drained mount in vertical position to prevent possible overpressure damage from frozen media 2 1 3 Static...

Page 5: ...t The device may only be installed and started up by specialist staff familiar with installation start up and operation of the product Specialist staff are defined as those individuals capable of asse...

Page 6: ...ce and pointer pursuant to EN 837 1 nominal size 100 mm Erl uter ung zus tzlich ei n Hi nweis 5 3 2 Instrument connection The instrument connection is located on the bottom side of the pressure gauge...

Page 7: ...sured the scope of the measurement range and of the protection against special conditions such as vibrations pulsations and pressure spikes A bracket must be installed to support the pressure gauge if...

Page 8: ...nduit that contains high current AC power cables Where possible avoid running the cable near inductive equipment Units with shielded cable electrical termination connect the drain wire to the guard te...

Page 9: ...secured against being switched on again 8 2 Check on function and recalibration Inspection of function and recalibration should be at regular intervals The precise testing cycles should be adjusted i...

Page 10: ...acement of the pressure gauge is required in the event any of the above exist 9 4 Conduct following fault rectification See Chapter 7 3 Mounting Installation 10 Removal disposal 10 1 Safety Residue fr...

Page 11: ...Page 11 of 11 11 2 CE Declaration of conformity...

Reviews: