Ashcroft F5509 Installation And Maintenance Instruction Manual Download Page 9

 

Page 9 of 27 

7.3  Storage 

The differential pressure gauge must be stored in dry, clean conditions, within a temperature range of -20 to +70 °C, 

protected against direct exposure to sunlight and protected against impact damage. 

8  Assembly/Installation 

8.1  Safety 

To ensure safe working during installation and servicing, suitable shut-off valves must be installed in the plant (see 6.4 

Accessories), enabling the device: 

 

To be depressurized or taken out of operation; 

 

To be disconnected from the pressure source for repair or check within the relevant plant; 

 

Or to enable function tests of the device to be performed “on site”. 

During the works to mount/install the gauge, the plant must be protected against being switched back on. 

8.2  Preparations (requirements for the installation location) 

 

A check on suitability of the device for the medium to be measured, the scope of the measurement range and 

static pressure and of the protection against special conditions such as vibration, pulsation and pressure spikes. 

 

A bracket must be installed to support the pressure gauge  if the mounting process pipe is not able to provide 

adequate support. 

 

The installation location should be chosen such that the work-spaces for operating personnel are not located to 

the rear of the pressure gauge. 

8.3  Mounting/Installation 

8.3.1  Process connection 

The instrument is intended and factory adjusted for vertical mounting, pressure ports downward. When mounted 

in other orientation (max. ± 10°) the pointers’ zero position needs to be adjusted (see 8.4.1 Zero point adjustment)

 

Connection to be made by authorized and qualified personnel only. 

 

Use only with the mechanical process connection provided – regarding the configuration, see order code on the 

device type label, with a matching threaded seal. 

 

When connecting the device, the process piping must be depressurized. 

 

The process pipe must be installed on an incline so that: 
- for fluid measurement, no air pockets are created 

- for gas measurement, no water pockets are created 

   If the necessary incline is not achieved, then at suitable points water separators or air separators must  be 

installed. 

 

The pressure  process  pipe  must  be kept as short as possible and installed  without sharp bends, to avoid a 

delayed response time. 

 

The instruments pressure ports are marked by "+" and "-" symbols:  

 

"+" port must be connected to the higher pressure  

 

"-" port must be connected to the lower pressure. 

 

With liquid measurement media, the pressurized connection pipe must be degassed, since any gas bubble 

inclusions result in measurement error. 

 

If water is used as the measurement medium, the device must be frost-protected. 

 
Safety notice: Only mount using the correct open-jawed wrench, and do not twist the device itself. 

 

8.3.2  Electrical connection 

 

Take note of the electrical data in the EU design type test certification and the locally-applicable regulations 

and guidelines for installing and operating electrical plant in explosion risk areas (e.g. EN 60079-14, 

EN 60079-17 etc.). 

Summary of Contents for F5509

Page 1: ...tenance Instruction Manual Differential pressure gauge model F5509 F6509 for industrial application in the following configuration F5509 F6509 ATEX differential pressure gauge without switching contac...

Page 2: ...nd F6509 without switching contact 5 3 2 F5509 I and F6509 I with inductive proximity switches SJ2 N 6 4 Technical data 7 5 Labeling on the device 7 5 1 Labeling on the device for explosion risk areas...

Page 3: ...t general safety information 2 and also the following specific advice regarding the intended purposes 2 2 and through to disposal 11 3 contain important safety information which if not followed may re...

Page 4: ...er will violate the approval for hazardous area and are not permitted This also applies to installation of spare parts Possible conversations or alterations may only be carried out by the manufacturer...

Page 5: ...clouds and dust layers the ignition temperature must be specified pursuant to ISO IEC 80079 20 2 For dust clouds the maximum allowable medium temperature inside of the measuring instrument must not ex...

Page 6: ...e in differential pressure between 10 and 90 of the measuring range and a pulse frequency of 0 06 Hz the temperature increase is 10 K To avoid additional temperature increase the devices should not be...

Page 7: ...tch operating voltage 230 VAC 4 Technical data The detailed technical information can be found in the documents in the Appendix Chapter 12 5 Labeling on the device The label with the serial number and...

Page 8: ...nnection Distance between ports is 37 mm please consider this when selecting a 5 way manifold 6 3 3 Vent valve The vent valve for the housing is located on the top side If the nipple is pulled out the...

Page 9: ...tical mounting pressure ports downward When mounted in other orientation max 10 the pointers zero position needs to be adjusted see 8 4 1 Zero point adjustment Connection to be made by authorized and...

Page 10: ...n site is possible For this proceed as follows Equalize pressure in both chambers Check if internal pressure was built up in the case due to ambient temperature effect Open valve position B wait for p...

Page 11: ...erformance and recalibration is carried out at regular intervals depending on the application The precise testing cycles should be adjusted in line with the operating conditions and ambient conditions...

Page 12: ...ty data sheets 11 2 Removal When servicing the gauge the process pressure line must be depressurized the electrical connections isolated from the mains supply and special attention should be taken to...

Page 13: ...Page 13 of 27 12 2 Declaration of conformity model F5509 F6509 without contacts...

Page 14: ...Page 14 of 27...

Page 15: ...Page 15 of 27 12 3 Declaration of conformity model F5509 and F6509 inductive proximity switches...

Page 16: ...Page 16 of 27...

Page 17: ...Page 17 of 27 12 4 EU design type test certification for inductive proximity switches of types SJ2 N...

Page 18: ...Page 18 of 27...

Page 19: ...Page 19 of 27...

Page 20: ...Page 20 of 27...

Page 21: ...Page 21 of 27...

Page 22: ...Page 22 of 27...

Page 23: ...Page 23 of 27...

Page 24: ...Page 24 of 27...

Page 25: ...Page 25 of 27...

Page 26: ...Page 26 of 27...

Page 27: ...Page 27 of 27...

Reviews: