Aseptico ADU-22 Operation And Maintenance Instruction Manual Download Page 6

2. HANDPIECE SELECTOR 

(Fig. 6) - The

Aseptico ADU-22 includes a handpiece

selector toggle on the front of the console

which selects either highspeed (HP1) or

lowspeed (HP2) handpiece operation. 
Connect the proper highspeed and

lowspeed handpieces to their appropriate

connector. Switch the handpiece selector

to the desired handpiece. Depress foot

control to activate handpiece.

The amount of air pressure supplied to

each handpiece may be adjusted by

turning the drive air-pressure control

screws underneath the console (See

page 7).

3. FOOT CONTROL 

(Fig. 7) - The foot

control applies air pressure to the selected

handpiece
Position the foot control on the floor. For

handpiece operation, apply foot pressure

to any part of the center disk. 

4

OPERATION FUNCTIONS:

1. THREE-WAY AIR/WATER SYRINGE 

-

Depress the right button for air operation,

and the left button for water operation.

Depressing both buttons will create a mist.

The syringe features quick-change

autoclavable tips: To remove a tip, press

on the locking collar surrounding the tip

socket and pull the used tip straight out of

the socket (see Fig.5). To insert a new tip,

press locking collar and push tip into

socket as far as it will go. Release collar

and gently tug on tip before using to

ensure that tip is securely locked into

socket.

Syringe Tip Sterilization:

1)

Remove contaminated syringe tip.

2)

Remove all visible signs of contamination

before autoclaving.

3)

Autoclave tip at 132° C (270° F) for ten

minutes.

4)

Sterilize between each patient use.

NOTE

: Since only the tips can be

autoclaved, it is recommended that the

air/water syringe be bagged with a

disposable, single-use plastic sleeve

between each patient use.

APPLY

PRESSURE TO

ACTIVATE

HANDPIECE

Figure 7 - Foot Control

Figure 6

LOWSPEED

CONTROL

HIGHSPEED

CONTROL

3-WAY

AIR/WATER

SYRINGE

HANDPIECE

SELECTOR

Figure 5

DEPRESS

LOCKING COLLAR

Summary of Contents for ADU-22

Page 1: ...OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU 22 DENTAL CART SYSTEM...

Page 2: ...4 Water Supply Tank 3 Operation Functions 4 Fig 5 TA 90D Air Water Syringe 4 Fig 6 Handpiece Controls 4 Fig 7 Foot Control 4 Fig 8 Control Console 5 Fig 9 Water Tank Removal 5 Fig 10 AA 21 HVE Vacuum...

Page 3: ...the need for a central vacuum pump Congratulations This Aseptico ADU 22 system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive t...

Page 4: ...IGH VOLUME VACUUM SYSTEM THREE WAY SYRINGE NWS 8 HDPE WATER SUPPLY TANK HIGHSPEED HANDPIECE CONTROL REMOVABLE INSTRUMENT TRAY 3 The ADU 22 unit must be connected to a dry filtered 60 90 psi air source...

Page 5: ...stall the HVE vacuum bottle press on locking ring tab Fig 3a push bottle into quick coupler and release tab 5b Hang the vacuum head in the holder at the right hand side of the chassis Fig 3b 6 To inst...

Page 6: ...ion and the left button for water operation Depressing both buttons will create a mist The syringe features quick change autoclavable tips To remove a tip press on the locking collar surrounding the t...

Page 7: ...emoving the tank for refilling it must be depressurized by turning the tank 1 4 turn counterclockwise The tank is removed by unscrewing it from its mount at the bottom right rear of the chassis After...

Page 8: ...e Fig 10 To operate remove the vacuum head from holder and insert a standard oral evacuator tip into end of head extension Depress the on off switch on the side of the hose end for vacuum Waste from t...

Page 9: ...ndpiece may be adjusted by turning the flow control screw for that handpiece The control screws are located underneath the console next to the HP1 and HP2 handpiece holders Figure 11 Handpiece Pressur...

Page 10: ...out 5 seconds after every patient and about 20 seconds at the beginning of each day 3 GENERAL CLEANING The cart exterior is designed to make cleaning and the use of prophylactic barriers as easy as po...

Page 11: ...9 NOTES...

Page 12: ...or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Cu...

Reviews: