Aseptico ADU-17A Operation And Maintenance Instruction Manual Download Page 8

8.

to decrease

 flow, or 

counterclockwise to increase

 flow.)

4.  Use the 3-Way Air/Water Syringe as necessary for irrigation 

or drying. 

5.  Adjust the vacuum pressure by moving the valves on both 

the HVE and saliva ejector vacuum heads. 

6.  Empty the Vacuum Waste Tank when it is 3/4 full.

STERILIZATION AND MAINTENANCE:

Because of its simple design, the Aseptico ADU-

17A Express Air requires very little maintenance. Any 

maintenance that is needed can be performed in minutes.

BLEEDING THE SYSTEM

If the unit will not be used for an extended period of time, 

or the unit might be subjected to freezing conditions, you 

should bleed the system. Simply empty the contents of the 

water supply tank and install the tank back into the cap. 

Operate the air/water syringe and handpiece with water 

coolant ‘ON’ until just air comes through the water line. 

Pack unit and store as normal.

HANDPIECE FLUSH

Flush the handpiece for about 5 seconds after every 

patient, and about 20 seconds at the beginning of each 

day. 

GENERAL CLEANING

The external surfaces of the case should be cleaned using 

a soft cloth moistened with a mild detergent solution. Any 

external surfaces of the unit that are contacted during use 

should be wiped down with a soft cloth moistened with 

a disinfectant at the beginning of each day and between 

each patient use.

WATER LINES

Disinfect the water lines weekly.  Prepare a 1:10 bleach 

solution (1 part household bleach to 9 parts water). 

Remove water reservoir and discard residual water. 

Replace empty water supply tank and air purge all 

waterlines. Fill water supply tank with bleach solution. Run 

bleach solution through all lines. Allow bleach solution to 

stand in lines for 10 minutes. Remove water supply tank 

and discard bleach. 

Flush water supply tank and all 

lines thoroughly with clean water.

 Air purge and leave 

lines dry until next clinical use.

VACUUM SYSTEM

The HVE and Saliva Ejector Valves are fully autoclavable. 

Remove the valves from the hoses before autoclaving. The 

vacuum hoses should not be autoclaved. Clean hoses with 

a disinfectant solution. CAUTION: Use only NON-foaming 

cleansers in the vaccum lines.

SYRINGE TIPS

The ADU-17A unit features a three-way air/water syringe 

with quick-change autoclavable tips. To remove a tip, press 

on the locking collar surrounding the tip socket and pull 

the used tip straight out of the socket (see Fig. 7). To insert 

a new tip, press locking collar and push tip into socket as 

far as it will go. Release collar and gently tug on tip before 

using to ensure that tip is securely locked into socket.

DEPRESS LOCKING 

COLLAR

FIG. 7

Syringe Tip Sterilization:

1.  Remove contaminated syringe tip.

2.  Remove all visible signs of contamination before 

autoclaving.

3.  Autoclave tip at 132° C (270° F) for ten minutes.

4.  Sterilize between each patient use.

NOTE:

 Since only the tips can be autoclaved, it is 

recommended that the Air/Water Syringe be bagged with a 

disposable, single-use plastic sleeve between each patient 

use.

This device requires periodic maintenance to remove 

dust in and around the compressor every 12 months or if 

accumulations are seen.  The fan on the unit vents air from 

inside the device to the outside.  With the unit operating, 

use compressed air to spray into the intake vents and blow 

dust out of the unit.

Summary of Contents for ADU-17A

Page 1: ...ADU 17A Express Air Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...sult if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60...

Page 3: ...st and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments may ca...

Page 4: ...dicator facing out Figure 2d IMPORTANT For 120V use check that the Fuse in use is a 6A medium time delay for 230V use check that the Fuse is a 4A Slo Blo type User may have to change fuses before use...

Page 5: ...ct the Low Volume Vacuum Saliva Ejector hose to the black connector located on the front of the Tank Fig 4 High Volume Connector Low Volume Connector Water Resevoir Connector Water Supply Bottle Waste...

Page 6: ...3 Way Air Water Syringe The Water Supply Bottle threads into the Reservoir Connector located in the upper right corner of the case To refill the Water Supply Bottle A Before removing the Bottle for re...

Page 7: ...nnect the power cord to the Express Air Unit and plug into appropriate power source IMPORTANT Ensure that the Voltage Selector Insert is properly configured for the power source Refer to Page 4 2 FOOT...

Page 8: ...all waterlines Fill water supply tank with bleach solution Run bleach solution through all lines Allow bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flu...

Page 9: ...heck handpiece connection for missing gasket Compressor makes a chugging sound and does not start up properly Check for sufficient power source and avoid long light grade extension cords Refer to Figu...

Page 10: ...ting 6A at 120 VAC or 4A at 230 VAC Power Rating 800 W Operating Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM...

Page 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Page 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420966 Rev G ECO 15498 04 2022...