5.

Fig. 3

Clip-On 

Holder

Handle

Saliva

Ejector

3-Way 

Syringe

Handpiece

HVE

4.  Attach a sterile handpiece to the ISO Type B (4-Line) 

handpiece connector and place the handpiece in Holder 

(Fig. 3).

5.  Remove the 3-way air/water Syringe from its travel clip 

inside the Case and place it in the Holder.

6.  Position the Waste Tank in the Case with the large white 

connection to the rear left of the instrument as shown 

in Figure 4. Attach the hose connector coming from the 

Vacuum Motor to this large white connection. Connect 

the High Volume Vacuum (HVE) hose to the large gray 

connector located on the front of the Waste Tank. Connect 

the Low Volume Vacuum (Saliva Ejector) hose to the 

black connector located on the front of the Tank.

 Fig. 4

High

Volume

Connector

Low

Volume

Connector

Water

Resevoir

Connector

Water

Supply

Bottle

Waste Tank

Vacuum

Motor

Connector

Bungee 

Cord

Note:

 When attaching the Saliva Ejector hose to the top of 

the Waste Tank, push in and rotate 1/4 turn (see Figure 5) 

7.  Secure Waste Tank by running Bungee Cord from hooks 

on bottom panel overtop Tank lid. Verify that the Waste 

Tank and vacuum lines are properly installed. Hang the 

High Vacuum and Saliva Ejector instruments on the Clip-

On Holder.

8.  Fill the Water Supply Bottle with clean water and connect 

to the Water Reservoir Connector.

9. 

Attach the foot pedal and place on the floor.

Voltage Selector Module

Fig. 2a

Fig. 2b

Fig. 2c

Fig. 2d

Fig. 2e

120V Indicator

240V Indicator

Summary of Contents for ADU-17A

Page 1: ...ADU 17A Express Air Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...sult if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60...

Page 3: ...st and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments may ca...

Page 4: ...dicator facing out Figure 2d IMPORTANT For 120V use check that the Fuse in use is a 6A medium time delay for 230V use check that the Fuse is a 4A Slo Blo type User may have to change fuses before use...

Page 5: ...ct the Low Volume Vacuum Saliva Ejector hose to the black connector located on the front of the Tank Fig 4 High Volume Connector Low Volume Connector Water Resevoir Connector Water Supply Bottle Waste...

Page 6: ...3 Way Air Water Syringe The Water Supply Bottle threads into the Reservoir Connector located in the upper right corner of the case To refill the Water Supply Bottle A Before removing the Bottle for re...

Page 7: ...nnect the power cord to the Express Air Unit and plug into appropriate power source IMPORTANT Ensure that the Voltage Selector Insert is properly configured for the power source Refer to Page 4 2 FOOT...

Page 8: ...all waterlines Fill water supply tank with bleach solution Run bleach solution through all lines Allow bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flu...

Page 9: ...heck handpiece connection for missing gasket Compressor makes a chugging sound and does not start up properly Check for sufficient power source and avoid long light grade extension cords Refer to Figu...

Page 10: ...ting 6A at 120 VAC or 4A at 230 VAC Power Rating 800 W Operating Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM...

Page 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Page 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420966 Rev G ECO 15498 04 2022...

Reviews: