TerraSpec® User Manual
Chapter 5
Batteries
The specifications of the battery are as follows:
To run the instrument with battery power, plug the battery cable from the
battery to the instrument power connector (back of instrument)
Note:
How long the spectrometer will run with a fully charged battery depends on
many variables including battery age, instrument configuration, environment,
and accessories powered by the accessory power port.
5.1
Battery Care
Proper care of the Gel Cell Electrolyte rechargeable batteries ensures a long
life.
The batteries are rated to give 200 to 300 uses when they are fully drained
(discharged to 9 Volts) each use. Letting the battery discharge under 9 Volts
will shorten the battery's life. If the battery is only used for 50% rated capacity
up to 500 uses are possible. In the same way using only 30% capacity can
result in up to 1,000 uses.
Note:
There is no such thing as memory effect on these cells. Only use them for the
time needed. After you're finished recharge them fully and then store away till
next use.
Type:
Gel Cell Electrolyte
Rating:
12 V 9 amp hour
Life:
Over four (4) hours using a full-range TerraSpec
spectrometer with a Contact Probe in an ambient
environment.
Nine (9) hours using a VNIR-only TerraSpec
spectrometer with accessory not powered by the
instrument in an ambient environment.
Recycle batteries and do not dispose of as general waste.
Summary of Contents for A100800
Page 1: ...TerraSpec User Manual ASD Document 600541 Rev G 2008 by ASD Inc www asdi com ...
Page 6: ...ASD Document 600541 Rev G vi TerraSpec User Manual www asdi com ...
Page 10: ...ASD Document 600541 Rev G x TerraSpec User Manual www asdi com Notes ...
Page 20: ...ASD Document 600541 Rev G 10 TerraSpec User Manual www asdi com Chapter 1 Introduction Notes ...
Page 44: ...ASD Document 600541 Rev G 34 TerraSpec User Manual www asdi com Chapter 5 Batteries Notes ...