INSTALLATION GUIDE
Card Reader (NUCR-HE)
3
Mounting
The NUCR has an IP40 ingress protection. The areas where NUCRs are to be installed must be clean, dry,
and weatherproof. The walls should be finished (painted, wall papered, tiled, etc.) before the devices are
installed.
The base of the NUCR mounts to backboxes using an array of holes that allow for mounting on various
backboxes.
Caution:
To prevent a fire hazard because of dust buildup inside the NUCR, regular
inspection and/or cleaning of the NUCR is required.
Take proper measures to avoid spreading dust and other particles when applying
maintenance to an NUCR that is mounted inside a clinical area.
3.1
Mount the NUCR onto a backbox
When Installing the NUCR with the tamper function active, a backbox must be used. The NUCR comes with
a special plastic plate for the tamper functionality.
Note that mounting the card reader on or near metal surfaces can negatively influence the RFID
card reader performance.
1.
For EU-style or UK-style backboxes, remove the screws from the backbox.
The screw distance for the backboxes must be 60mm (2.36 in.).
2.
Place the plastic plate over the backbox with the arrow pointing up (item 1).
The protrusion must point forward (item 2).
3.
Pull the room buses through the plastic plate of the NUCR.
3
TD 93429EN / 18 March 2022 / Ver. C