4
Charge the battery
Note:
For ATEX/IECEx transceivers, see ATEX/IECEx safety
instructions.
An empty battery icon in the display and a beep signal played
once every 10 s. during one minute together with a fast flashing
red LED shows a low battery. Place the transmitter in a single
charger or a wall mounted charging rack. If a charging rack is
used for charging, just insert the transmitter without opening the
clip, to make it fasten securely. Make sure to hear a beep and see
the LED charging indication when inserting the transmitter in the
charging rack.
A flashing battery icon and orange LED indicates charging. A
green steady LED and a filled battery icon shows a fully charged
battery. This takes approximately two hours.
Note:
Check the battery status every time you start to use the
transmitter. It is recommended that the transmitter is charged
daily. This is even more important when the transmitter is used in
temperatures below 0o C.
Note:
While charging, Man-down & No-movement alarms are
automatically deactivated.
Test alarm
Press and hold the Alarm push-button until Test Alarm is
displayed.
Note:
It is recommended to send a test alarm every day to test
the transceiver and the system.
Personal alarm
Press the Alarm push-button twice within two seconds.
Personal Alarm is displayed.
Pull-cord alarm
An alarm is sent when the Pull-cord is pulled off the transceiver.
The Pull-cord is also used as a security string.
Note:
Make sure that the clip of the pull-cord is safely attached
to your clothing. To make a test, remove the battery of the
transceiver, pull the transceiver and make sure that the pull-cord
releases first.
Alarm functions