background image

 

A. 

Introduction

B. Warnings&Cautions

1. Warning:

2. Cautions

Infrared  thermometer   should  be protected   for the

following

1). EMF  (electro-magnetic    fields)  from  arc welders,
induction  heaters.
2). Thermal  shock  (cause  by large  or abrupt  ambient
temperature   changes  allow  30 minutes  for  unit to
stabilize  before  use).
3). Do not  leave  the unit  on or near  objects  of high
temperature.

1. When take measurement, pay attention to the Distance
to Spot Size. As the Distance (D) from the target surface
increases, the spot size (S) of the area measured by the
unit becomes larger.
The Distance to Spot size of the unit is 12:1.
This unit is equipped with a laser, which is used for aiming.

2. Field of view:

Make sure the target is larger than the unit's spot size.
The smaller the target the closer measure distance.
When accuracy is critical, make sure the target is at least
twice as large as the spot size.

C. Distance to spot size

38mm@
300mm

75mm@
900mm

132mm@
1500mm

1.5"@
12"

3.0"@
36"

5.3"@
60"

D:S=12:1

Figure. 1

Emissivity:   Most organic   materials  and  painted  or
oxidized  surfaces   have an  emissivity  of 0.95(preset   in
the unit).  Inaccurate   readings  will  result  from  measuring
shiny  or polished  metal  surfaces.   To compensate  for  this,
adjust  the units  emissivity   reading  or cover  the surface   
to

 

be measured   with masking  tape  or flat  black  paint.

Measure  the tape  or painted  surface   when the  tape or
painted  reach  the same  temperature   as the material
underneath.

Trigger

Figure 3

E. Operation

D. Emissivity

1.  

Operating the unit

1). Open the battery door and insert a 

      

9V battery properly.

2). Pull the trigger to turn on the unit
3). Aim at the target surface and pull 

     

the trigger, then temperature will 

     

be shown on the LCD.

     

This unit is equipped with a laser, 

     

which is only used for

 

aiming.  

    

2. Locating a Hot Spot

To find a hot spot, aim the thermometer outside of interest,
then scan across with an up and down motion until you
locate the hot spot. (Figure 

3

)

 

 

Figure. 2

and processing circuit. 

This infrared thermometer is used for measuring the
temperature of  the object's surface, which is applicable 

for various hot, hazardous or hard-to reach objects

without contact safely and quickly. 

This unit consist of Optics, Temperature Sensor Signal 
amplifier, Processing circuit and LCD Display. The Optics
collected the infrared energy emitted by object and focus
onto the Sensor. Then the sensor translates the energy

into an electricity signal. This signal will be turned out 
to be digital shown on the LCD after the signal amplifier

To avoid the potential situation may cause harm or dam-
age to people, please pay attention to the following items:
1) . Do not point laser directly at eye or indirectly off reflec-
tive surfaces.
2) . The unit cannot measure through transparent surfaces
such as glass or plastic. It will measure the surface tem-

perature of these materials instead.

3) . Steam, dust, smoke, or other particles can prevent acc-

urate measurement by obstructing by the units optics

as - Schwabe GmbH   ·   Hummelbergstr. 6   ·   72184 Eutingen   ·   www.as-schwabe.de

Material

Emissivity

Material

Emissivity

Aluminum

Asbestos

Asphalt

Basalt

Brass

Brick

Carbon

Ceramic

Concrete

Copper

Dirt

Frozen  food

Hot food

Glass(plate)

Ice

Iron

Lead

Limestone

Oil

Paint

Paper

Plastic

Rubber

Sand

Skin

Snow

Steel

Textiles

Water

Wood

0.30

0.95

0.95

0.70

0.50

0.90

0.85

0.95

0.95

0.95

0.94

0.90

0.93

0.85

0.98

0.70

0.50

0.98

0.94

0.93

0.95

0.95

0.95

0.90

0.98

0.90

0.80

0.94

0.93

0.94

F. LCD Display & buttons

1. LCD display: Abb. 4

A: measuring reading
B: measuring unit
C: laseron icon
D: back light on icon
E: battery power icon
F: scanning icon
G: data  hold icon
H: mode/emissivity indicator
I: data  storge  / read icon
J: low temperature alarm icon
K: high temperature alarm icon

(1) Trigger:   When turn  on LCD display  VERXX 

   

software

 

version  for 1 sec.  And turn  to display  

   

reading  with

 

SCAN icon.  Release  the  trigger,  

   

display  reading  with

 

HOLD icon.  Built  in auto  

   

power  off in 7sec.

(2) Laser  / back  light button:   when back  light  

   

turn on ,

 

any operations   will  remain  back light  

   

for 10  sec. LCD

 indicate  on/off  status.

(3)—(6) key functions: press key

(3)

, LCD subdisplay

   

blinks MAX-MIN-DIF-AVG-HAL-LAL-STO-EMS seg-

   

ment(only main display means normal measuring

   

mode)  press key 

(4)

 to enter.

a. MAX:  measuring  maximum  temperature
b. MIN:  measuring  minimum  temperature
c. DIF:  Basic on  the reading  before  press key

 (4)

,

   compute the difference of current  reading.

d. AVG: measuring average temperature
e. HAL:high temperature alarm--when selected HAL,

   

press keys

 (5)

 to set  high temperature  alarm 

   

trigger

 

and confirmed   by pressing  key

(4)

.  

   

When reading over

 trigger,  LCD display HI icon 

   with BiBi  audio  sounds.

f. LAL:  low temperature alarm--when selected LAL,

   

press 5 keys to set  low temperature alarm 

t

rigger

   

and

 

confirmed by pressing key

(4)

.

   

When reading over trigger, LCD  display LOW icon

   

with BiBi  audio sounds

.

g. STO: data storage-when selected STO, lock&DATA

   

& 1--indicator  will shown when press key

(4)

. After

   

temperature read

 

out press 6 key to store, then 2--

   

memory unit will be shown. There 30 groups 

   

memory

 

unit available. To recall the stored data 

   

in normal

 

measuring mode by

 

pressing  key

(6)

   

remove all data

 

by pressing key

(6)

 for 3 seconds.

H. EMS: Emissivity setup-

 

press key 

(5)

 for emissivity

    

settings, press  key

(4)

  to save setup  and  back to

   normal  status.

(7) Clesius /Fahrenheit  switch:  Please open battery

    and push the slide  switch  for convertsion.

1. 

1. Lens Cleaning: 

Blow off loose particles using clean compressed

 

air. Gently brush remaining debris away with a 
moist cotton swab. The swab may be moistened

with water.

2. Case cleaning: Clean the case with a damp 
spo

nge/cloth and mild soap.

Note:
      1) Do not use solvent to clean plastic lens.
      2) Do not submerge the unit in water.

G. Maintenence

H. Specification

Accuracy

0

~900

 (32

~1652

±

1.5

(

±

2.7

)

or

±

1.5%

-50

~0

(-58

~32

)

 

±

3

 (

±

5

)

(whichever is greater)

Repeatability

1% of reading or 1

Response time

500 mSec, 95% response 

8-14 um

Emissivity

0.1 - 1.0 /  0.95 preset 

0 ~40

 (32 ~ 104

)

-20 ~ 60

 (-4~140

Power

9V Alkaline  or  NiCd battery

Non-laser   mode: 22  hrs;
Laser  Models:12  hrs

12:1

Spectral response

Operating Humidity

Storage Temperature

Typical battery life

(Alkaline)

Distance to Spot size

  -50 ~ 900°C

 (

-58 ~ 1652°F

)

Temperature range

Operating Temperature

10~95%RH  non-condensing

up to 30

(86

Resolution

0.1

 or 0.1

Figure 5

Figure. 6

2. Diagram description

(figure 

5

)