3.5.3 Quad 8 to 1 Multiplexer Mode (20-745 Only)
First, to check the size of the multiplexer at address 21.
TYPE? 9
will return “
MUX(4[8])
”.
To close channel 7 (bank 4) on multiplexer 9.
CHAN 9,4,7
Now to look at status of multiplexer 9, bank 4.
VIEW? 9,4
will return “
7
”.
To clear bank 4.
ARESET 9,4
Now to look at status.
VIEW? 9,4
will return “
0
", the status of bank 4.
VIEW? 9
will return the status of all multiplexer banks.