Cleaning and Sterilization
Surface Cleaning & Disinfection of Controller and Foot Control
DO NOT
sterilize or immerse in liquid. Wipe clean with a soft cloth and mild detergent as needed.
Clean with detergents and disinfectants according to standard practices. Disinfect with liquid chemical
disinfectants such as chlorine solutions, iodophors, glutaraldehydes and hydrogen peroxides. Follow
manufacturer guidelines for concentration and length of exposure.
Cleaning and Sterilization of Patient Cable
The Patient Cable is supplied NON-STERILE. It is reusable if cleaned and sterilized prior to each use.
Ensure that a validated and approved steam sterilization process is used during the sterilization
process.
Surface Cleaning and Disinfection
Before cleaning, secure the protective caps over the Patient Cable connectors to protect them from
possible damage. The protective caps should fit snugly over the Patient Cable connector. If the
protective caps are loose, check to see if the silicone o-ring is positioned within the groove on the
cable connector. Replacement o-rings are available through ArthroCare. The Patient Cable should
then be wiped down with a mild, disinfecting detergent solution using a soft brush or sponge to
remove any gross contaminants from the Patient Cable. The Patient Cable can be rinsed under
running water to remove any cleaning residue. The Patient Cable should not be immersed in water
or any other solution. Remove the protective caps from the Patient Cable connectors and inspect
the Patient Cable for any damage such as cuts or nicks. Damaged Patient Cable should not be
reused. Sterilize the clean Patient Cable by following one of the recommended sterilization
methods:
Pre-Vacuum Steam Sterilization
If the pre-
vacuum steam sterilizer has a prefixed cycle, use the “Hard Goods” cycle. If it does not have
a prefixed cycle, the following parameters are recommended:
Set temperature at 270-272°F (132-133°C);
Set exposure time for 10 minutes for wrapped Patient Cables;
Set exposure time for 4 minutes for unwrapped Patient Cables;
Set drying time for 5 minutes minimum.
Gravity Displacement Steam Sterilization
The following parameters are recommended for gravity displacement steam sterilization:
Set temperature at 270-272°F (132-133°C);
Set exposure time to 15 minutes for wrapped or unwrapped Cables;
Set drying time for 8 minutes minimum.
Or
Set temperature at 250-254°F (121-123°C);
Set exposure time to 30 minutes for wrapped or unwrapped cables;
Set drying time for 8 minutes.
CAUTION:
Failure to properly clean the Patient Cable may lead to inadequate sterilization.