DFU-0320-3 Rev. 0
17 of 20
Cleaning
The following cleaning instructions are provided.
CAUTIONS
ABRASION -
Do not use abrasive cloth, sharp objects, or abrasive cleaners.
DISCONNECTION -
Detach the interconnect cables and the AC adapter from
the receiver.
IMMERSION
- Do not immerse the receiver, cables or connectors under
running water.
Transmitter
The transmitter is
IP68 rated
and
can be
completely immersed briefly.
Dampen a non-abrasive cloth with one of the following products; then wring out
until slightly wet and gently rub soiled area until clean.
Isopropyl alcohol
Soap and water
Cidex
Sodium Hypochlorite 5.25% (Bleach) diluted 10:1
Receiver
CAUTION
: The receiver is
IPX1
rated and
cannot
be immersed.
Dampen a non-abrasive cloth with one of the following products; then wring out
until slightly wet and gently rub soiled area until clean.
Isopropyl alcohol
Soap and water
Cidex
Bleach Sodium Hypochlorite 5.25% (Bleach) diluted 10:1
Wipe any fluids from the surface of the receiver.