DrillSaw Mini 300™ System
User’s Guide
DFU-0223-2 Rev. 0
Page 3
1.
DO NOT
– under any condition, or for any reason – open the handpiece, any
system component, or any other system accessory.
2. This device has passed testing for EMI/RFI radiation and susceptibility, and
EMC compatibility. This device may cause interference with other devices in the
near vicinity if not set up, and used as instructed by Arthrex.
3. Do not attach accessories to the handpiece unless the handpiece is in the safe
mode.
4. Only use
batteries approved by Arthrex.
5.
Do NOT
clean the device with abrasive cleaning or disinfectant compounds,
solvents, or other materials that could scratch or damage the device.
6. The battery packs are
NOT
to be cleaned and disinfected in a thermo washer
disinfector.
7.
NEVER
place the handpiece or accessories in Cidex or other aldehyde
disinfectant solutions.
8. After sterilization in the autoclave, let the accessory device cool down slowly.
NEVER
use cold water to cool the handpieces. This will damage the electronic
components and seals.
9. Liquid on the battery and handpiece contacts may damage the device. Before
connecting the battery housing, ensure the receptacles are clean and dry.
10. Always comply with the instructions issued by the manufacturer of the cleaning
disinfectant regarding concentration, exposure times, temperature, and material
compatibility.
11. Remove all accessories, including attachments, battery pack, and accessories
before sterilizing a handpiece.