DrillSaw Mini 300™ System
User’s Guide
DFU-0223-2 Rev. 0
Page 24
4.11
Battery Housing/Battery Pack/Transfer Plate
These instructions describe the procedure for inserting the battery into the battery
housing from the AR-300.
Figure 13
Insert drop-in battery to the Battery Housing
Table 18
Drop-in battery
A
Sterile Technician
B
Non-sterile Circulating Nurse
1.
The Sterile Technician holds sterile battery housing upright and
places the sterile transfer plate on top.
2.
The Non-sterile Circulating Nurse places the unsterile battery down into
the housing through the sterile transfer plate until it is fully seated.
3.
The Non-sterile Circulating Nurse then removes the sterile
transfer plate, and the Sterile Technician closes the sterile battery
housing door.
3a - The assembled battery is now ready to be placed on the
handpiece.